[How reliable is a hot-wire anemometer?].

H von Rechenberg, H Konder, K Höser

    Der Anaesthesist
    |August 1, 1985
    PubMed
    Summary
    This summary is machine-generated.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Expanding the role of mammographers: a training strategy to enhance mammographer-patient interaction.

    Breast disease·2005
    Same author

    ARDS in fulminant ornithosis and treatment with extracorporeal lung assist.

    The International journal of artificial organs·2000
    Same author

    [Intra- and postoperative airway management of tracheal carina ruptures from bilateral endobronchial intubation].

    Anasthesiologie, Intensivmedizin, Notfallmedizin, Schmerztherapie : AINS·1999
    Same author

    [Rett's syndrome: pathophysiology and anesthesiological implications].

    Anasthesiologie, Intensivmedizin, Notfallmedizin, Schmerztherapie : AINS·1999
    Same author

    [Blood gas analysis in interhospital transfer--a useful extension of respiratory monitoring?].

    Anasthesiologie, Intensivmedizin, Notfallmedizin, Schmerztherapie : AINS·1999
    Same author

    Elimination of drugs by the new polyamide hemofilter FH77H during various in vitro conditions.

    Blood purification·1998

    Hot-wire anemometers require careful validation for clinical use. Electronic noise suppression and transducer changes can distort measurements, limiting reliable clinical application.

    Area of Science:

    • Biomedical Engineering
    • Clinical Measurement
    • Respiratory Physiology

    Context:

    • Hot-wire anemometry is utilized for airflow measurement in clinical settings.
    • Assessing the reliability and validity of hot-wire anemometers is crucial for accurate patient monitoring.
    • Previous studies have not fully addressed the limitations of these devices in clinical environments.

    Purpose:

    • To evaluate the clinical utility of hot-wire anemometers.
    • To identify sources of measurement error in hot-wire anemometry.
    • To establish criteria for reliable clinical measurements using hot-wire anemometers.

    Summary:

    • Two types of hot-wire anemometers were tested for reliability and validity in a clinical context.
    • Electronic noise suppression was found to distort measurements.

    Related Experiment Videos

  • Transducer variations also led to deviations from true airflow values.
  • Impact:

    • Manufacturers should provide device perception thresholds and coefficients of variation.
    • A method is proposed to estimate reliable measurement limits based on equipment thresholds and ventilation parameters.
    • Improved understanding of hot-wire anemometer limitations can enhance patient respiratory monitoring accuracy.