Comparison of Iodine Density Measurement Among Dual-Energy Computed Tomography Scanners From 3 Vendors
Objectives:
The aims of this study were to analyze the effect of dual-energy computed tomography (DECT) scanners and fluid characteristics on iodine quantification and to calculate the measurement variability range induced by those variables.
Methods:
We performed an experimental phantom study with 4 mediastinal iodine phantoms. Each phantom contained 6 tubes of different iodine concentrations (0, 1.0, 2.5, 5.0, 10.0, and 20.0 mg/mL) diluted in a specific solvent, which was water, 10% amino acid solution, 20% lipid emulsion, or 18% calcium solution, respectively. Mediastinal phantoms were inserted into an anthropomorphic chest phantom and were scanned with 3 different DECT scanners from 3 vendors using 2 radiation dosage settings. Iodine density (IoD) and computed tomography (CT) attenuation at virtual monoenergetic 70-keV images and virtual nonenhanced images were measured for the iodine phantoms. The effects of DECT scanners, solvents, and radiation dosage on the absolute measurement error of IoD and on the CT attenuation profiles were investigated using linear mixed-effects models. Measurement variability range of IoD was also determined.
Results:
Absolute error of IoD was not significantly affected by the DECT systems, kind of solvents, and radiation dosage settings. However, CT attenuation profiles were significantly different among the DECT vendors and simulated body fluids. Measurement variability range of IoD was from -0.6 to 0.4 mg/mL for the true iodine concentration 0 mg/mL.
Conclusions:
Dual-energy CT systems and fluid characteristics did not have a significant effect on the IoD measurement accuracy. A cutoff of IoD for the determination of a truly enhancing lesion on DECT would be 0.4 mg/mL.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Strain-Energy Density
In the elastic region of a material, the relationship between the stress and the strain is linear and follows Hooke's Law. The strain energy density in this region...
Imaging Studies III: Computed Tomography
Molecular Comparison of Gases, Liquids, and Solids
Cell Potential and Free Energy
Thermodynamics is the branch of physics dealing with the relationship between heat and other forms of energy. In an electrochemical cell, chemical energy is converted into electrical energy.
Thus, a link can be predicted between cell potential, free energy change, and the equilibrium constant for the reaction. Cell potential can also be measured as the oxidant or the reducing strength, and similar acid-base strength measures are reflected in equilibrium...
Free Energy