Aims:
The RELAX-OAB study is designed to confirm the safety, efficacy, and technical performance of the Axonics r-SNM System, a miniaturized, rechargeable SNM system approved in Europe and Canada for the treatment of bladder and bowel dysfunction. The purpose of this article is to describe study subjects' ability to charge the rechargeable neurostimulator and to document their neurostimulator program settings and recharge interval over time.
Methods:
Fifty-one OAB patients were implanted in a single-stage procedure. These results represent the 3-month charging experience for 48 subjects who completed the 3-month follow-up. Recharge intervals were estimated using therapy stimulation settings and subject experience was evaluated using questionnaires.
Results:
Forty-seven of forty-eight (98%) subjects were able to successfully charge their device prior to follow-up within 1-month post-implant. At 3-month post-implant, 98% of subjects were able to charge prior to their follow-up visit. Average stimulation amplitude across all subjects was 1.8 mA (±1.1 mA). A total of 69% of subjects had ≥14-day recharge intervals (time between charging) and 98% of subjects had ≥7-day recharge interval. No charging related adverse events occurred.
Conclusions:
Study subjects were able to charge the Axonics r-SNM System and stimulation settings provided 2 weeks of therapy between recharging for most subjects. Subject satisfaction indicates that subjects are satisfied with rechargeable SNM therapy.
Related Concept Videos
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Prediction Intervals
However, the point estimate is most likely not the exact value of the population parameter, but close to it. After calculating point estimates, we construct interval estimates, called confidence intervals or prediction intervals. This prediction interval comprises a range of values unlike the point estimate and is a better predictor of the observed sample value, y.
Confidence Intervals
A...
Improper Integrals: Infinite Intervals
Urinary Bladder
In males, the bladder is situated in front of the rectum, while in females, it is positioned anterior to the vagina and uterus. The bladder floor contains an inverted triangular area called the trigone, defined by the two ureteric...
Interval Level of Measurement
Data measured using the interval scale are similar to ordinal level data because they have a definite arrangement. However, in the interval level of measurement, the differences between data values are meaningful even though the data does not have a starting point.
Temperature is measured using the interval scale. It is measurable data, and the difference between...