Comparison of maximal Rubidium-82 activities for myocardial blood flow quantification between digital and
Joris D van Dijk1, Pieter L Jager2, Jochen A C van Osch3
1Department of Nuclear Medicine, Isala hospital, PO Box 10400, 8000 GK, Zwolle, The Netherlands. jorisvdijk@gmail.com.
Background:
PET-based myocardial blood flow (MBF) quantification can be inaccurate when using high tracer activities. Our aim was to derive the maximal Rubidium-82 activity for MBF assessment using a new digital PET system and compare the results with conventional analog systems.
Methods:
1.8 GBq Rubidium-82 was injected into the cardiac insert of an anthropomorphic torso phantom. Data were acquired for 10 min using an Ingenuity TF (Philips Healthcare), Discovery 690 (D690, GE Healthcare), and digital PET prototype system (Philips Healthcare). The dynamic ranges, defined as the maximal measured activity in the reconstructed images deviating < 10% from the true present activity, were determined in all scans.
Results:
The dynamic ranges were 312 MBq for Ingenuity TF, 650 MBq for D690, and 654 MBq for digital PET prototype.
Conclusions:
The maximal Rb-82 activity for MBF assessment using digital PET prototype is higher than that for its analog counterpart (Ingenuity TF), but seems comparable to the D690.
Related Concept Videos
Blood Flow
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
Applications of Integration to Find Blood Flow
Sign Convention
The normal force acts perpendicular to the beam's cross-section and can...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Enolate Mechanism Conventions