Related Experiment Video
Updated: Feb 15, 2026

Comparative Analysis of Automatic Fecal Analyzer versus Direct Wet Smear Microscopy for Detecting Parasitic Infections in Stool Samples
Published on: April 25, 2025
Low cost automated whole smear microscopy screening system for detection of acid fast bacilli
Yan Nei Law1, Hanbin Jian1, Norman W S Lo2
1Hong Kong Applied Science & Technology Research Institute Co., Ltd. (ASTRI), Hong Kong SAR, China.
Background:
In countries with high tuberculosis (TB) burden, there is urgent need for rapid, large-scale screening to detect smear-positive patients. We developed a computer-aided whole smear screening system that focuses in real-time, captures images and provides diagnostic grading, for both bright-field and fluorescence microscopy for detection of acid-fast-bacilli (AFB) from respiratory specimens.
Objectives:
To evaluate the performance of dual-mode screening system in AFB diagnostic algorithms on concentrated smears with auramine O (AO) staining, as well as direct smears with AO and Ziehl-Neelsen (ZN) staining, using mycobacterial culture results as gold standard.
Methods:
Adult patient sputum samples requesting for M. tuberculosis cultures were divided into three batches for staining: direct AO-stained, direct ZN-stained and concentrated smears AO-stained. All slides were graded by an experienced microscopist, in parallel with the automated whole smear screening system. Sensitivity and specificity of a TB diagnostic algorithm in using the screening system alone, and in combination with a microscopist, were evaluated.
Results:
Of 488 direct AO-stained smears, 228 were culture positive. These yielded a sensitivity of 81.6% and specificity of 74.2%. Of 334 direct smears with ZN staining, 142 were culture positive, which gave a sensitivity of 70.4% and specificity of 76.6%. Of 505 concentrated smears with AO staining, 250 were culture positive, giving a sensitivity of 86.4% and specificity of 71.0%. To further improve performance, machine grading was confirmed by manual smear grading when the number of AFBs detected fell within an uncertainty range. These combined results gave significant improvement in specificity (AO-direct:85.4%; ZN-direct:85.4%; AO-concentrated:92.5%) and slight improvement in sensitivity while requiring only limited manual workload.
Conclusion:
Our system achieved high sensitivity without substantially compromising specificity when compared to culture results. Significant improvement in specificity was obtained when uncertain results were confirmed by manual smear grading. This approach had potential to substantially reduce workload of microscopists in high burden countries.
Related Concept Videos
Sputum Studies I: Gram Stain, cytology, and Acid-fast smear and culture
Gram Stain
The Gram Stain is an integral part of sputum studies. It involves the staining of sputum, which permits...
Fast Fourier Transform
The computational efficiency of the FFT becomes...
Fast Decoupled and DC Powerflow
Gene Evolution - Fast or Slow?
In contrast, regions which code...
Gene Evolution - Fast or Slow?
Amino acids