Related Concept Videos

Pulse Oximetry01:24

Pulse Oximetry

1.4K
Pulse oximetry, or SpO2, is a non-invasive method for continuously monitoring arterial oxygen saturation (SaO2). This procedure involves attaching a probe or sensor to the patient's fingertip, forehead, earlobe, or nose bridge. The sensor works by detecting changes in oxygen saturation levels through light signals generated by the oximeter and reflected by the pulsing blood under the probe.
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
1.4K
Reliability and Validity01:29

Reliability and Validity

14.1K
Reliability and validity are two important considerations that must be made with any type of data collection. Reliability refers to the ability to consistently produce a given result. In the context of psychological research, this would mean that any instruments or tools used to collect data do so in consistent, reproducible ways.
14.1K
Pulse01:16

Pulse

2.2K
When the heart pumps blood out, arterial elastic fibers play a crucial role in sustaining a high-pressure gradient. They expand to accommodate the received blood and then recoil - a process known as the pulse that can be either manually palpated or electronically quantified. Despite a reduction in its effect with increased distance from the heart, elements of the pulse's systolic and diastolic components persist, observable even at the arteriole level.
The pulse serves as a clinical...
2.2K
Pulse01:05

Pulse

4.2K
The pulse is one of the most fundamental physiological indicators of the body's cardiovascular health. It is the rhythmic expansion and contraction of the arterial walls in response to the pressure generated by the heart's pumping action.
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
4.2K
Theoretical Approaches to Psychological Disorder01:29

Theoretical Approaches to Psychological Disorder

885
The development of psychological disorders, which are characterized by deviant, maladaptive, and personally distressing behaviors, has been explored through several theoretical approaches.
Biological approach
The biological approach posits that internal, organic factors are the primary causes of such disorders. This perspective emphasizes brain structure and function, genetic predispositions, and neurotransmitter imbalances. For example, schizophrenia has been associated with both genetic...
885
Theoretical Foundations of Nursing Practice01:30

Theoretical Foundations of Nursing Practice

18.0K
Theories play an essential role in organizing patient care. Theories refer to a proposed or followed belief, policy, or procedure that is the basis for action. Nursing theories are knowledge-based concepts that guide nurses' actions, influence nursing education and practice, and allow nurses to care for their patients.
Theories provide a perspective to assess patients' conditions and organize data and methods. They also assist in analyzing and interpreting information. They represent a...
18.0K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Clinical validation of video-based vital sign monitoring in the intensive care unit: a prospective cohort study : Clinical validation of video-based vital sign monitoring.

Journal of clinical monitoring and computing·2026
Same author

Thermal Cameras for Overnight Measuring of Respiration in a Clinical Setting.

Sensors (Basel, Switzerland)·2025
Same author

Accuracy of remote, video-based supraventricular tachycardia detection in patients undergoing elective electrical cardioversion: a prospective cohort.

Journal of clinical monitoring and computing·2025
Same author

Thermal Cameras for Continuous and Contactless Respiration Monitoring.

Sensors (Basel, Switzerland)·2025
Same author

Speckle Vibrometry for Contactless Instantaneous Heart Rate and Respiration Rate Monitoring on Mechanically Ventilated Patients.

Sensors (Basel, Switzerland)·2024
Same author

Speckle Vibrometry for Instantaneous Heart Rate Monitoring.

Sensors (Basel, Switzerland)·2023

Related Experiment Video

Updated: Feb 15, 2026

An R-Based Landscape Validation of a Competing Risk Model
05:37

An R-Based Landscape Validation of a Competing Risk Model

Published on: September 16, 2022

2.6K

Camera-based pulse-oximetry - validated risks and opportunities from theoretical analysis.

Mark Van Gastel1, Sander Stuijk1, Gerard De Haan1,2

  • 1Department of Electrical Engineering, Eindhoven University of Technology, PO Box 513, 5600MB, Eindhoven, The Netherlands.

Biomedical Optics Express
|January 24, 2018
PubMed
Summary

Optimizing camera-based pulse oximetry with three wavelengths enhances measurement accuracy. This study identifies optimal wavelengths and a reliability index, improving robustness against noise and dyshemoglobins.

Keywords:
(170.0170) Medical optics and biotechnology(170.1470) Blood or tissue constituent monitoring(170.3660) Light propagation in tissues(280.0280) Remote sensing and sensors

More Related Videos

Imaging and Quantification of the Area of Fast-Moving Microbubbles Using a High-Speed Camera and Image Analysis
05:31

Imaging and Quantification of the Area of Fast-Moving Microbubbles Using a High-Speed Camera and Image Analysis

Published on: September 5, 2020

6.4K
Author Spotlight: Assessing the Cardiovascular Profile of Patients with Metabolic Syndrome
06:04

Author Spotlight: Assessing the Cardiovascular Profile of Patients with Metabolic Syndrome

Published on: September 27, 2024

1.6K

Related Experiment Videos

Last Updated: Feb 15, 2026

An R-Based Landscape Validation of a Competing Risk Model
05:37

An R-Based Landscape Validation of a Competing Risk Model

Published on: September 16, 2022

2.6K
Imaging and Quantification of the Area of Fast-Moving Microbubbles Using a High-Speed Camera and Image Analysis
05:31

Imaging and Quantification of the Area of Fast-Moving Microbubbles Using a High-Speed Camera and Image Analysis

Published on: September 5, 2020

6.4K
Author Spotlight: Assessing the Cardiovascular Profile of Patients with Metabolic Syndrome
06:04

Author Spotlight: Assessing the Cardiovascular Profile of Patients with Metabolic Syndrome

Published on: September 27, 2024

1.6K

Area of Science:

  • Biomedical Engineering
  • Optical Sensing

Background:

  • Camera-based pulse oximetry shows promise but is susceptible to noise and motion artifacts.
  • Existing methods using two wavelengths lack optimal selection for improved robustness.
  • Factors like dyshemoglobins can cause inaccurate oxygen saturation (SpO2) readings.

Purpose of the Study:

  • To identify optimal wavelengths for camera-based pulse oximetry to enhance measurement robustness.
  • To investigate the impact of physiological factors on pulse oximetry measurements using a theoretical skin model.
  • To develop a reliability index for pulse oximetry and detect/quantify dyshemoglobins.

Main Methods:

  • Performed a systematic search for optimal wavelengths in camera-based pulse oximetry.
  • Utilized a theoretical skin model to analyze the influence of various factors on SpO2 measurements.
  • Compared estimated carboxyhemoglobin (COHb) levels in smokers and non-smokers.

Main Results:

  • Identified optimal wavelengths that significantly improve measurement robustness against noise and artifacts.
  • Demonstrated that a third wavelength can create a reliable index for SpO2 measurements.
  • Successfully detected and quantified dyshemoglobins, showing differences in COHb levels between smokers and non-smokers.

Conclusions:

  • Optimal wavelength selection is crucial for robust camera-based pulse oximetry.
  • A multi-wavelength approach enhances accuracy and provides a reliability metric.
  • Camera-based pulse oximetry can detect and quantify dyshemoglobins, offering insights into physiological conditions.