Negative nucleotide ions as sensitive probes for energy specificity in collision-induced fragmentation in mass
Stefanie Ickert1,2, Jens Riedel2, Sebastian Beck1
1Department of Chemistry, Humboldt-Universitaet zu Berlin, Brook-Taylor-Str. 2, 12489, Berlin, Germany.
Rationale:
The most commonly used fragmentation methods in tandem mass spectrometry (MS/MS) are collision-induced dissociation (CID) and higher energy collisional dissociation (HCD). While in CID the preselected ions in the trap are resonantly (and m/z exclusively) excited, in HCD the entire m/z range experiences the dissociative acceleration. The different excitation is reflected in different fragment distributions.
Methods:
As a test-bed for particularly pronounced fragmentation specificity, here MS/MS experiments on several 4-mer oligonucleotides were conducted employing both collision methods and the results were thoroughly compared. Oligonucleotides are shown to be sensitive probes to subtle changes, especially in the negative ion mode. A detailed analysis of these differences reveals insight into the dissociation mechanics.
Results:
The differences are represented in heat-maps, which allow for a direct visual inspection of large amounts of data. In these false colour representations the, sometimes subtle, changes in the individual dissociation product distributions become distinct. Another advantage of these graphic plots can be found in the formation of systematic patterns. These patterns reflect trends in dissociation specificity which allow for the formulation of general rules in fragmentation behavior.
Conclusions:
Instruments equipped with two different excitation schemes for MS/MS are today widely available. Nonetheless, direct comparisons between the individual results are scarcely made. Such comparative studies bear a powerful analytical potential to elucidate fragmentation reaction mechanism.
Related Concept Videos
Mass Spectrometry: Alkene Fragmentation
Mass Spectrometry: Amine Fragmentation
In amines, the number of nitrogen atoms affects the mass of the molecular ion, which is described by the nitrogen rule of mass spectrometry. This rule states that a compound containing a single...
Mass Spectrometry: Cycloalkane Fragmentation
For example, cyclohexane molecular ions have a mass-to-charge ratio (m/z) of 84, which tends to produce a stronger signal than linear alkanes like hexane. This stability comes from...
Mass Spectrometry: Cycloalkene Fragmentation
Mass Spectrometry: Alkyne Fragmentation
Mass Spectrometry: Alcohol Fragmentation