Injecting without pressing a button: An exploratory study of a shield-triggered injection mechanism
Eric Zijlstra1, Hans-Veit Coester1, Tim Heise1
1Profil, Neuss, Germany.
Aims:
To evaluate the injection success and user perception of a shield-triggered pen-injector mechanism.
Methods:
The trial (ClinicalTrials.gov NCT02627287) was an exploratory, two-centre, one-visit, open-label, randomized controlled trial conducted in Germany in 150 injection-experienced individuals with type 1 or type 2 diabetes. Participants self-administered subcutaneous injections of a placebo solution using a prototype shield-triggered pen-injector, DV3316 (Novo Nordisk, Bagsvaerd, Denmark), and FlexPen (Novo Nordisk, Bagsvaerd, Denmark). Injection success was evaluated on a yes/no basis by the investigator. Participant confidence, leakage of fluid and pain were evaluated after each injection. Pain and device experience were assessed after completion of all injections with each pen-injector. Overall preference was assessed after completion of all injections with both pen-injectors.
Results:
Injection success was high with both pen-injectors (97.0%, DV3316 vs 99.7%, FlexPen). Participant confidence in dose delivery was similar for the two devices (88% of injections with DV3316 vs 81% with FlexPen were scored as "extremely confident"). The median injection pain score on a visual analogue scale (0-100) was 3 with DV3316 vs 4 with FlexPen after each injection, and 4 with DV3316 vs 5 with FlexPen after all injections with each device. After all injections were completed, 55% of participants reported an overall preference for DV3316 vs 21% for FlexPen.
Conclusion:
This study demonstrates that injection-experienced individuals can achieve a high injection success rate with a shield-triggered pen-injector, with similar patient confidence and injection pain compared with FlexPen.
Related Concept Videos
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Nonlinear Pharmacokinetics: Drug Elimination for IV Bolus Injection
Following the administration of a single intravenous (IV) bolus injection, we can determine the concentration of the drug in the plasma at any given time. This calculation is achieved using a specific equation that integrates the values of Vmax and KM.
We can also...
Drug Accumulation During Multiple Dosing: Repetitive IV Injections
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Reaction Mechanisms
For instance, the decomposition of ozone appears to follow a mechanism with two steps: