Biomechanical comparisons of current suspensory fixation devices for anterior cruciate ligament reconstruction
Cheng Jin1, Siddhartha Venkata Paluvadi1,2, SungJae Lee3
1Center for Joint Diseases, Chonnam National Univerity Bitgoeul Hospital, Chonnam National University School of Medicine, 160 Ilsimri, Hwasun-eup, Hwaun-gun, 519-809, Hwasun, Jeolla-Namdo, South Korea.
Purpose:
Cortical suspensory devices are routinely used for femoral side fixation of soft tissue graft in anterior cruciate ligament (ACL) reconstruction. The purpose of this study was to evaluate the biomechanical properties of a new adjustable loop device (GraftMax®) compared with established devices (EndoButton® and TightRope®) in ACL reconstruction and to investigate whether knotting the free end of TightRope could improve biomechanical properties.
Methods:
The three cortical suspensory devices (GraftMax® Button; Conmed, EndoButton® CL; Smith & Nephew, and TightRope® RT; Arthrex) were tested under cyclic load (50-250 N for 1000 cycles) and pull-to-failure conditions at 50 mm/h in a device-only setup using a tensile testing machine. The TightRope was additionally tested with its free suture ends knotted. The statistical analyses were done with one-way analysis of variance (ANOVA) and post hoc Tukey HSD tests.
Results:
There are significant differences in the load-to-failure among the devices. The EndoButton showed the highest mean failure load at 1204.7 N compared to other devices (GraftMax (914.2 N), knotted TightRope (868.1 N) and TightRope (800.1 N) (p < 0.001). The mean total displacement after 1000 cycles was 0.76 mm, 2.11 mm, 1.56 mm and 1.38 mm for the EndoButton, GraftMax, TightRope, and knotted TightRope, respectively. The EndoButton showed significantly better properties than both the GraftMax (p = 0.000) and the TightRope (p = 0.020) but not the knotted TightRope (p = n.s.) in total displacement. However, there was no significant difference between the TightRope and GraftMax (p = n.s.).
Conclusion:
The fixed loop (EndoButton) showed significantly better mechanical properties in failure load and displacement than TightRope or GraftMax in this biomechanical study. However, the mechanical properties of the GraftMax is comparable to the TightRope. Moreover, the knotting of TightRope improved mechanical properties in total displacement more than TightRope, but not in failure load.
Clinical Relevance:
The biomechanical properties of the GraftMax are comparable to the TightRope. The TightRope, when knotted, shows an improvement both in load to failure and cyclic displacement, though the differences are not significant.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Muscles of the Anterior Neck
Carbon-dioxide Fixation
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Electrical Current