A Practical Guide on Coupling a Scanning Mobility Sizer and Inductively Coupled Plasma Mass Spectrometer SMPS-ICPMS
Published on: July 11, 2017
Concentration-Gradient-Method for sulphur and strontium isotope ratio determination by quadrupole-based inductively
Michail Dronov1,2, Torsten C Schmidt2, Jürgen Schram1
1University of Duisburg-Essen, Instrumental Analytical Chemistry and Centre for Environmental and Water Research (ZWU), Universitätsstr. 5, D-45141, Essen, Germany.
Rationale:
The concentration-gradient-method (CGM) was previously introduced as a precise and accurate method for isotope ratio determination by quadrupole-based inductively coupled plasma mass spectrometry (ICP-QMS). The investigation of its potential and advantages in the analysis of analytes with a poor signal-noise ratio (S/N) is important to establish routine isotope ratio analysis industrial applications on these widely used instruments.
Methods:
The CGM was applied on isotope measurements of Sr near its limit of quantification (LOQ) and of sulphur where there was a massively interfered 32 S isotope signal in gypsum samples of different origin, in order to demonstrate the advantages of the CGM over the commonly used measurement and evaluation approach. The comparison between the CGM and the classical measurement and evaluation approach was performed with high and low concentration Sr standard solutions, to prove the robustness of the isotope ratio determination.
Results:
In both cases the CGM reached a recovery rate of approximately 103 %, whereas the classical approach became increasingly inaccurate at lower S/N (recovery of 123 %). In the case of sulphur isotope ratio determination only the CGM enabled a differentiation between geogenic and flue gas desulphurisation plant-originated gypsum samples.
Conclusions:
The robustness of the CGM approach was illustrated for gypsum samples with trace strontium concentration and its general applicability to the determination of sulphur isotope ratios by means of quadrupole-based ICP-MS was demonstrated using the example of sulphur in gypsum.
More Related Videos
07:25Quantitative Metabolomics of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: January 5, 2021
09:05Imaging Metals in Brain Tissue by Laser Ablation - Inductively Coupled Plasma - Mass Spectrometry LA-ICP-MS
Published on: January 22, 2017
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometry: Isotope Effect
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Isotopes
An element's atomic mass, or weight,...