Assessment of Physical Activity Intensity with Accelerometers and Oxygen Consumption
Published on: June 20, 2025
Comparison of Two Generations of ActiGraph Accelerometers: The CARDIA Study
Kara M Whitaker1, Kelley Pettee Gabriel2, David R Jacobs3
1Division of Health and Human Physiology, College of Liberal Arts and Sciences, University of Iowa, Iowa City, IA.
Purpose:
This study aimed to examine the comparability of the ActiGraph 7164 and wGT3X-BT wear time, count-based estimates, and average time per day in physical activity of different intensities.
Methods:
We studied 87 Coronary Artery Risk Development in Young Adults (CARDIA) participants 48-60 yr of age who simultaneously wore the 7164 and wGT3X-BT accelerometers at the waist in 2015-2016, with wear time of ≥4 of 7 d, ≥10 h·d for both monitors. Freedson cutpoints (counts per minute) were used to define sedentary (<100), light (100-1951), moderate (1952-5724), and vigorous activity (≥5725). Agreement was evaluated using paired-difference tests, intraclass correlation coefficients, and Bland-Altman plots. Given systematic differences in count-based estimates between monitors, a calibration formula applied to the wGT3X-BT values was obtained by linear regression.
Results:
Total detected wear time minutes per day was nearly identical between the 7164 and the wGT3X-BT (881.5 ± 70.9 vs 880.3 ± 78.1, P = 0.72). The wGT3X-BT values were calibrated to the 7164 values by dividing counts by 1.088. After calibration, no differences were observed between the 7164 and the wGT3X-BT in total counts per day (310,184 ± 129,189 vs 307,085 ± 135,362, P = 0.48), average counts per min per day (349.5 ± 139.5 vs 346.5 ± 147.2, P = 0.54), sedentary (513.2 ± 93.6 vs 509.6 ± 98.6, P = 0.23), light (335.3 ± 81.5 vs 338.7 ± 81.1, P = 0.22), moderate (31.0 ± 21.9 vs 30.3 ± 23.4, P = 0.31), or moderate-to-vigorous minutes per day (33.1 ± 24.6 vs 32.0 ± 26.0, P = 0.13). A significant difference was observed for vigorous minutes per day (0.2 ± 1.0 vs 0.0 ± 0.3, P < 0.01); however, the absolute difference was marginal. Intraclass correlation coefficients showed excellent agreement for all measures (0.95-0.99).
Conclusions:
After applying a calibration formula, the 7164 and wGT3X-BT were comparable for total wear time, count-based estimates, and average minutes per day in sedentary, light, moderate, and moderate-to-vigorous activity. Findings illustrate a novel methodological approach to facilitate accelerometer data harmonization.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits
Molecular Comparison of Gases, Liquids, and Solids