Related Experiment Video
Updated: Feb 15, 2026

Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Accelerated real-time cardiac MRI using iterative sparse SENSE reconstruction: comparing performance in patients with
Bradley D Allen1, Maria L Carr2, Michael Markl3,4
1Department of Radiology, Northwestern University, 737 N. Michigan Ave, Suite 1600, Chicago, IL, 60611, USA. bdallen@northwestern.edu.
Objectives:
To compare accelerated real-time cardiac MRI (CMR) using sparse spatial and temporal undersampling and non-linear iterative SENSE reconstruction (RT IS SENSE) with real-time CMR (RT) and segmented CMR (SEG) in a cohort that included atrial fibrillation (AF) patients.
Methods:
We evaluated 27 subjects, including 11 AF patients, by acquiring steady-state free precession cine images covering the left ventricle (LV) at 1.5 T with SEG (acceleration factor 2, TR 42 ms, 1.8 × 1.8 × 6 mm3), RT (acceleration factor 3, TR 62 ms, 3.0 × 3.0 × 7 mm3), and RT IS SENSE (acceleration factor 9.9-12, TR 42 ms, 2.0 × 2.0 × 7 mm3). We performed quantitative LV functional analysis in sinus rhythm (SR) patients and qualitatively scored image quality, noise and artefact using a 5-point Likert scale in the complete cohort and AF and SR subgroups.
Results:
There was no difference between LV functional parameters between acquisitions in SR patients. RT IS SENSE short-axis image quality was superior to SEG (4.5 ± 0.6 vs. 3.9 ± 1.1, p = 0.007) and RT (3.8 ± 0.4, p = 0.003). There was reduced artefact in RT IS SENSE compared to SEG (4.4 ± 0.6 vs. 3.8 ± 1.2, p = 0.04), driven by arrhythmia performance. RT IS SENSE short-axis image quality was superior to SEG (4.6 ± 0.5 vs. 3.1 ± 1.0, p < 0.001) in the AF subgroup.
Conclusion:
Accelerated real-time CMR with iterative sparse SENSE provides excellent clinical performance, especially in patients with AF.
Key Points:
• Iterative sparse SENSE significantly accelerates real-time cardiovascular MRI acquisitions. • It provides excellent qualitative and quantitative performance in sinus rhythm patients. • It outperforms standard segmented acquisitions in patients with atrial fibrillation. • It improves the trade-off between temporal and spatial resolution in real-time imaging.
Related Concept Videos
Electrophysiology of Normal Cardiac Rhythm
Pathophysiology of Cardiac Performance
Amyloid Fibrils
Amyloid deposits were observed as early as 1639 in the liver and the spleen. In 1854, Rudolph Virchow performed iodine staining,...
Amyloid Fibrils
Real Time RT-PCR
The real-time quantification of the number of amplified products is...
The Sense of Self: Reflected Self-Appraisal and Social Comparison