Transcorporal Artificial Urinary Sphincter Cuff Placement in a Case Requiring Revision for Urethral Atrophy
Published on: June 16, 2022
Montreal electronic artificial urinary sphincters: Our futuristic alternatives to the AMS800™
Xavier Biardeau1, Sami Hached2, Oleg Loutochin1
1Jewish General Hospital, McGill University; Montreal, QC, Canada.
Introduction:
We aimed to present three novel remotely controlled hydromechanical artificial urinary sphincters (AUSs) and report their in-vitro and ex-vivo results.
Methods:
We successively developed three distinct hydromechanical AUSs on the basis of the existing AMS800™ device by incorporating an electronic pump. No changes were made to the cuff and balloon. The AUS#1 was designed as an electromagnetically controlled device. The AUS#2 and AUS#3 were conceived as Bluetooth 2.1 remotely controlled and Bluetooth 4.0 remotely-controlled, adaptive devices, respectively. In-vitro experiments profiled occlusive cuff pressure (OCP) during a complete device cycle, with different predetermined OCP. Ex-vivo experiments were performed on a fresh pig bladder with 4 cm cuff placed around the urethra. Leak point pressure with different predetermined OCP values was successively measured during cystometry via a catheter at the bladder dome.
Results:
Our in-vitro and ex-vivo experiments demonstrated that these three novel AUSs provided stable and predetermined OCP - within the physiological range - and completely deflated the cuff, when required, in a limited time compatible with physiological voiding cycles.
Conclusions:
Our three novel, remotely controlled AUSs showed promising results that should be confirmed by in-vivo experiments focusing on efficacy and safety.
More Related Videos
10:06Engineering Artificial Factors to Specifically Manipulate Alternative Splicing in Human Cells
Published on: April 26, 2017
10:07Cystometric and External Urethral Sphincter Measurements in Awake Rats with Implanted Catheter and Electrodes Allowing for Repeated Measurements
Published on: January 30, 2018
Related Concept Videos
Alternative RNA Splicing
There are five types of alternative RNA splicing that vary in the ways the pre-mRNA segments are removed or retained in the mature mRNA. The first...
Alternative RNA Splicing
Null and Alternative Hypotheses
The null hypothesis, denoted by H0 is a statement of no difference between the variables—they are not related. This can often be considered the status quo. As a result if you cannot accept the null, it requires some action.
The alternative hypothesis, denoted by H1 or Ha, is a claim about the...
Electron Carriers
Over the many stages of cellular respiration, glucose breaks down into carbon dioxide and water. Electron carriers pick up electrons lost by glucose in these reactions, temporarily storing and releasing them into the electron...
Introduction to Urinary System
The kidneys are bean-shaped organs located in the retroperitoneal space, on either side of the vertebral column, between the T12 and L3 vertebrae. They are partially protected by the rib cage and surrounded by perirenal fat, which provides cushioning. They are responsible for urine formation and play critical roles in regulating blood pressure, electrolyte levels, and hormone production. The ureters...
Urinary Bladder
In males, the bladder is situated in front of the rectum, while in females, it is positioned anterior to the vagina and uterus. The bladder floor contains an inverted triangular area called the trigone, defined by the two ureteric...