Accuracy of Corneal Thickness by Swept-Source Optical Coherence Tomography and Scheimpflug Camera in Virgin and
Francisco Arnalich-Montiel1, Sara Ortiz-Toquero2, Clara Auladell1
1Department of Ophthalmology, Ramón y Cajal University Hospital, Madrid, Spain.
Purpose:
To assess intraobserver repeatability, intersession reproducibility, and agreement of swept-source Fourier-domain optical coherence tomography (SS-OCT) and the Scheimpflug camera in measuring corneal thickness in virgin and grafted eyes with Fuchs endothelial corneal dystrophy (FECD).
Methods:
Thirty-six control eyes, 35 FECD eyes, 30 FECD with corneal edema eyes, 25 Descemet stripping automated endothelial keratoplasty (DSAEK) eyes, and 29 Descemet membrane endothelial keratoplasty (DMEK) eyes were included. The apical center, pupillary center, and thinnest corneal thickness were determined in 3 consecutive images and repeated 2 weeks later. Repeatability and reproducibility coefficients, intraclass correlation coefficients, and 95% limits of agreement (LOA) between measurements were calculated. Agreement between devices was assessed using Bland-Altman analysis.
Results:
Corneal thickness measurements were highly reproducible and repeatable with both systems. SS-OCT showed better repeatability in all corneal locations in the normal, FECD, FECD with edema, DSAEK, and DMEK groups (coefficient of variation ≤0.60%, ≤0.36%, ≤0.43%, ≤1.09%, and ≤0.48%, respectively) than the Scheimpflug (coefficient of variation ≤1.15%, ≤0.92%, ≤1.10%, ≤1.25%, and ≤1.14%, respectively). Between-session 95% LOA for SS-OCT was less than 3% for all groups except for the FECD with edema group, being almost double using the Scheimpflug camera. Differences between instruments were statistically significant in all groups and locations (P < 0.01) except in the DSAEK group (P ≤ 0.51); however, SS-OCT underestimated all measurements.
Conclusions:
SS-OCT provides better reproducible and repeatable measurements of corneal thickness than those obtained with the Scheimpflug camera in patients with FECD or an endothelial transplant. Variations between examinations higher than the 95% LOA observed in our study should raise awareness of changes in the endothelial function.
More Related Videos
03:35Determining Gender-Based Differences in Retinal and Choroidal Thickness in Underweight Individuals via Swept-Source Optical Coherence Tomography
Published on: December 1, 2023
07:51Full-Field Optical Coherence Microscopy for Histology-Like Analysis of Stromal Features in Corneal Grafts
Published on: October 21, 2022
Related Concept Videos
Improving Translational Accuracy
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Accuracy and Precision
Satellite Stem Cells and Muscular Dystrophy
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...