GRASE Revisited: breath-hold three-dimensional (3D) magnetic resonance cholangiopancreatography using a Gradient and
Ju Gang Nam1,2, Jeong Min Lee1,2,3, Hyo-Jin Kang1,2
1Department of Radiology and Institute of Radiation Medicine, Seoul National University Hospital, 101 Daehak-ro, Jongno-gu, Seoul, 03080, Republic of Korea.
Objective:
To evaluate the clinical feasibility and image quality of breath-hold (BH) three-dimensional (3D) magnetic resonance cholangiopancreatography (MRCP) using a gradient and spin-echo (GRASE) technique compared to the conventional 3D respiratory-triggered (RT)-MRCP using a turbo spin-echo (TSE) sequence at 3 T.
Methods:
Sixty-six patients underwent both 3D RT-TSE-MRCP and 3D BH-GRASE-MRCP at 3 T. Three radiologists independently reviewed the visualisation of biliary and pancreatic ducts, image blurring, and overall image quality of the two data sets using four- or five-point scales. The numbers of scans with non-diagnostic or poor image quality were compared between the two scans.
Results:
The 3D BH-GRASE-MRCP had a significantly better image quality (3.69 ± 0.77 vs. 3.30 ± 1.18, p = 0.005) and less image blurring (3.23 ± 0.94 vs. 3.65 ± 0.57, p = 0.0003) than the 3D RT-TSE-MRCP. In detail, 3D BH-GRASE-MRCP better depicted the common bile duct, cystic duct, and bilateral first intrahepatic duct (all ps < 0.05). The number of scans with non-diagnostic or poor image quality significantly decreased with 3D BH-GRASE-MRCP compared with 3D RT-TSE-MRCP [19.7% (13/66) vs. 1.5% (1/66), p = 0.002].
Conclusion:
The 3D BH-GRASE-MRCP provided better image quality and a reduced number of non-diagnostic images compared to 3D RT-TSE-MRCP.
Key Points:
• The GRASE technique enabled 3D MRCP acquisition within a single breath-hold. • The short acquisition time of 3D BH-GRASE-MRCP significantly reduced image blurring. • The 3D BH-GRASE-MRCP had a better image quality than 3D RT-TSE-MRCP. • The number of non-diagnostic scans was reduced with 3D BH-GRASE-MRCP.
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Magnetic Resonance Imaging
Breathing
Atomic Nuclei: Magnetic Resonance
Nuclear Magnetic Resonance (NMR): Overview
NMR spectroscopy generates a spectrum where the characteristic absorption frequencies of the sample are...
What is an Electrochemical Gradient?
The chemical gradient relies on differences in the abundance of a substance on the outside versus the inside of a cell and flows from areas of high to low ion concentration. In contrast, the electrical gradient revolves around an...