Assessment of patient's peak skin dose during abdominopelvic embolization using radiochromic (Gafchromic) films
J Greffier1, J Goupil1, A Larbi1
1EA 2415, department of radiology, Nîmes university hospital, medical imaging group Nîmes, CHU de Nîmes, boulevard Prof. Robert-Debré, 30029 Nîmes cedex, France.
Purpose:
To assess the value of the routine use of radiochromic films in abdominopelvic embolization procedures to improve patient follow-up.
Methods:
A total of 55 patients who underwent transcatheter abdominopelvic embolization were prospectively included. Six types of procedures were evaluated including hepatic chemoembolization (HCE), gonadal veins embolization (GVE), uterine elective embolization (UEE), uterine urgent embolization (UUE), abdominal elective embolization (AEE), and abdominal urgent embolization (AUE). Dosimetric indicators (DIs) such as air-kerma (AK) and kerma-area-product (KAP) were collected and peak skin dose (PSD) was measured with radiochromic films. Correlations between PSD and DIs were searched for.
Results:
The mean (±standard deviation [SD]) PSD for the various procedures were: 1033±502 mGy for HCE; 476±271 mGy for GVE; 460±171 mGy for UEE; 531±263 mGy for UUE; 708±896 mGy for AEE; 683±392 mGy for AUE. Strong correlations were observed between PSD and DIs (r=0.974 for AK and r=0.925 for KAP). PSD was>2Gy in one procedure and all procedures (7/132) procedures resulted in AK>2Gy, mostly for HCE and AEE.
Conclusion:
Dosimetry using radiochromic film is only appropriate for HCE, AEE and AUE, whereas dose-mapping systems present a more suitable solution for all embolizations including those with AK that occasionally exceed 2Gy.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography