Feasibility of using an inversion-recovery ultrashort echo time (UTE) sequence for quantification of glenoid bone
Ya-Jun Ma1,2, Justin West3, Amin Nazaran1,2
1Department of Radiology, University of California, San Diego Medical Center, San Diego, CA, 92103, USA.
Objective:
To utilize the 3D inversion recovery prepared ultrashort echo time with cones readout (IR-UTE-Cones) MRI technique for direct imaging of lamellar bone with comparison to the gold standard of computed tomography (CT).
Materials And Methods:
CT and MRI was performed on 11 shoulder specimens and three patients. Five specimens had imaging performed before and after glenoid fracture (osteotomy). 2D and 3D volume-rendered CT images were reconstructed and conventional T1-weighted and 3D IR-UTE-Cones MRI techniques were performed. Glenoid widths and defects were independently measured by two readers using the circle method. Measurements were compared with those made from 3D CT datasets. Paired-sample Student's t tests and intraclass correlation coefficients were performed. In addition, 2D CT and 3D IR-UTE-Cones MRI datasets were linearly registered, digitally overlaid, and compared in consensus by these two readers.
Results:
Compared with the reference standard (3D CT), glenoid bone diameter measurements made on 2D CT and 3D IR-UTE-Cones were not significantly different for either reader, whereas T1-weighted images underestimated the diameter (mean difference of 0.18 cm, p = 0.003 and 0.16 cm, p = 0.022 for readers 1 and 2, respectively). However, mean margin of error for measuring glenoid bone loss was small for all modalities (range, 1.46-3.92%). All measured ICCs were near perfect. Digitally registered 2D CT and 3D IR-UTE-Cones MRI datasets yielded essentially perfect congruity between the two modalities.
Conclusions:
The 3D IR-UTE-Cones MRI technique selectively visualizes lamellar bone, produces similar contrast to 2D CT imaging, and compares favorably to measurements made using 2D and 3D CT.
More Related Videos
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Line Loss
Line loss impacts power delivery efficiency in a balanced three-phase circuit. The symmetry in such a circuit simplifies the...
Inverse Trigonometric Functions
Reducing Line Loss
With a step-up transformer at the source, the voltage is increased, thereby reducing the current in the transmission lines since power loss in...
Inverse Hyperbolic Functions and Their Derivatives
Derivatives of Inverse Trigonometric Functions