Image Quality on Dual-energy CTPA Virtual Monoenergetic Images: Quantitative and Qualitative Assessment
Bari Dane1, Hersh Patel1, Thomas O'Donnell2
1Department of Radiology, NYU Langone Health, 660 First Avenue, New York, NY 10016.
Rationale And Objectives:
This study aims to determine the optimal photon energy for image quality of the pulmonary arteries (PAs) on dual-energy computed tomography (CT) pulmonary angiography (CTPA) utilizing low volumes of iodinated contrast.
Materials And Methods:
The study received institutional review board exemption and was Health Insurance Portability and Accountability Act compliant. Adults (n = 56) who underwent dual-energy CTPA with 50-60 cc of iodinated contrast on a third-generation dual-source multidetector CT were retrospectively and consecutively identified. Twelve virtual monoenergetic kiloelectron volt (keV) image data sets (40-150 keV, 10-keV increments) were generated with a second-generation noise-reducing algorithm. Standard regions of interest were placed on main, right, left, and right interlobar pulmonary arteries; pectoralis muscle; and extrathoracic air. Attenuation [mean CT number (Hounsfield unit, HU)], noise [standard deviation (HU)], signal to noise (SNR), and contrast to noise ratio were evaluated. Three blinded chest radiologists rated (from 1 to 5, with 5 being the best) randomized monoenergetic and weighted-average images for attenuation and noise. P <.05 was considered significant.
Results:
Region of interest mean CT number increased as keV decreased, with 40 keV having the highest value (P < .001). Mean SNR was highest for 40-60 keV (P <.05) (14.5-14.7) and was higher (P <.05) than all remaining energies (90-150 keV) for all vessel regions combined. Contrast to noise ratio was highest for 40 keV (P <.001) and decreased as keV increased. SNR was highest at 60 and 70 keV, only slightly higher than 40-50 keV (P <.05). Reader scores for 40-50 keV were greater than other energies and weighted-average images (P <.05).
Conclusions:
Kiloelectron volt images of 40-50 keV from the second-generation algorithm optimize attenuation on dual-energy CTPA and can potentially aid in interpretation and avoiding nondiagnostic examinations.
More Related Videos
15:10A Dual Tracer PET-MRI Protocol for the Quantitative Measure of Regional Brain Energy Substrates Uptake in the Rat
Published on: December 28, 2013
07:24A Hybrid DNA Extraction Method for the Qualitative and Quantitative Assessment of Bacterial Communities from Poultry Production Samples
Published on: December 10, 2014
Related Concept Videos
Qualitative Analysis
For instance, group IV...
Qualitative Analysis
There are two main approaches to qualitative analysis:...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Virtual Work
In static equilibrium, a body can experience an imaginary or virtual movement, such as displacement or rotation. The virtual work done by a force is equal to the dot product of force and virtual displacement in the direction of the force. When it comes to virtually rotating a...
What is Energy?
Free Energy