Techniques for trans-catheter retrieval of embolized Nit-Occlud® PDA-R and ASD-R devices
Sanjay Sinha1, Daniel Levi1, Alejandro Peirone2
1Division of Cardiology, Department of Pediatrics, UCLA Mattel Children's Hospital, Los Angeles, California.
Background:
Nit-Occlud® (atrial septal defect) ASD-R and (patent ductus arteriosus) PDA-R devices are used outside the United States for percutaneous closure of the patent ductus arteriosus and atrial septal defects. When embolization occurs, these devices have been difficult to retrieve.
Methods:
Bench simulations of retrieval of PDA-R and ASD-R devices were performed in a vascular model. Retrieval of each device was attempted using snare techniques or with bioptome forceps with a range of devices. The same devices were then intentionally embolized in an animal model. Retrieval methods were systematically tested in a range of sheath sizes, and graded in terms of difficulty and retrieval time.
Results:
Devices that were grasped by the bioptome in the center of the proximal part of the devices were easily retrieved in both models. Bench studies determined the minimum sheath sizes needed for retrieval of each device with this method. In general sheathes two french sizes greater than the delivery sheath were successful with this technique. Three out of the four PDA-R devices were successfully retrieved in vivo. Two were retrieved by grasping the middle of the PA end of the PDA-R device with a Maslanka bioptome and one small PDA-R device was retrieved using a 10 mm Snare. Four of the five ASD-R devices were retrieved successfully grasping the right atrial ASD-R disc or by passing a wire through the device and snaring this loop. For ASD-R 28 and 30 mm devices, a double bioptome technique was needed to retrieve the device.
Conclusion:
ASD-R and PDA-R devices can be successfully retrieved in the catheterization lab. It is critical to grab the center portion of the right atrial disc of the ASD-R device or pulmonary portion of the PDA-R device and to use adequately sized sheathes.
More Related Videos
Related Concept Videos
Retrieval
Recall involves accessing information without cues, such as during an essay test, where individuals must retrieve facts and concepts from memory unaided. Another example is remembering the name of a colleague...
Pulmonary Embolism I: Introduction
ER Retrieval Pathway
The ER uses many checkpoints to prevent the entry of incorrectly folded or a resident protein as cargo onto a transport vesicle. These mechanisms...
Pulmonary Embolism III: Nursing Management
Disubstituted Cyclohexanes: cis-trans Isomerism
In cyclohexane, the substituents can occupy different positions generating distinct isomers....
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care