Motion effects on the measurement of stiffness on ultrasound shear wave elastography: a moving liver fibrosis phantom
Hyun Joo Shin1, Myung-Joon Kim2, Choon-Sik Yoon3
1Department of Radiology and Research Institute of Radiological Science, Severance Children's Hospital, Yonsei University College of Medicine, Seoul, Korea. lamer-22@yuhs.ac.
Aims:
To evaluate the differences between shear wave velocities (SWVs) measured with ultrasound elastography during the continuous motion using liver fibrosis phantoms.
Materials And Methods:
Elasticities were measured with convex and linear transducers of supersonic shear wave imaging (SSI) and acoustic radiation force impulse imaging (ARFI) using liver elasticity phantoms (3.0 and 16.9 kPa) at depths of 2, 3, 4, and 5 cm. Motion velocities were 30 and 60 rpm with the phantoms in an upright position on the Orbital shaker. To simulate different directional motion, the phantoms were laid on their side on the shaker. The values between moving and static status were compared, and the number of measurement failure was counted. Results: In SSI, the convex transducer was less affected by motion at 30 rpm with the 3 kPa phantom. In the higher velocity motion and in the higher stiffness phantom, most values from SSI were different comparing with static status, and there was a tendency for elasticity values to increase during movement. In ARFI, there were frequent measurement failures without stable results during the motion.
Conclusions:
Motion affected the measurement of elasticity differently in SSI and ARFI, according to the velocity, direction of the motion, and phantom stiffness. The convex transducer of SSI was less affected by motion in lower velocity motion and when using normal liver stiffness phantom.
More Related Videos
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
06:09Author Spotlight: Advancing Hepatic Fibrosis Diagnosis Using Magnetic Resonance Elastography and AI
Published on: July 21, 2023
Related Concept Videos
Equations of Wave Motion
The Wave Nature of Light
Shear Diagram
First, a free-body diagram of the beam is drawn, representing all the external forces and internal reactions acting on the beam. One can calculate the reaction forces at each support by employing the equilibrium equations of force and moment. The vertical component...
Shearing Stress
The average shearing stress can be calculated by dividing the shear by the area of the cross-section.
Shearing Strain
Cystic Fibrosis: Pathogenesis
CF is primarily caused by a genetic mutation in a chromosome 7 gene coding for the cystic fibrosis transmembrane conductance regulator (CFTR) protein. The most common gene mutation leading to CF is the ΔF508 mutation,...