Investigations of organ and effective doses of abdominal cone-beam computed tomography during transarterial
Yi-Shuan Hwang1,2, Hui-Yu Tsai3, Yu-Ying Lin2
1Department of Medical Imaging and Intervention, Chang Gung Memorial Hospital at Linkou, 5 Fushing Street, Kweishan, Taoyuan, 333, Taiwan, Republic of China.
Background:
To investigate the organ dose, effective dose (ED), conversion factor, and the C-arm rotation angle effects on dose variations of abdominal C-arm cone-beam computed tomography (CBCT) during transarterial chemoembolization (TACE).
Methods:
The organ doses and EDs for abdominal C-arm CBCT were retrospectively calculated according to a Monte Carlo technique for 80 patients. Dose variations from projections, ED to dose-area product (DAP) ratios, and effects of body mass index (BMI) on the ED and ED to DAP ratios were also analyzed.
Results:
The kidney received the highest dose (14.6 ± 1.2 mSv). Organ dose deviations among C-arm rotation angles was highest for stomach (CV = 0.71). The mean ED of the the CBCT run during TACE was 3.5 ± 0.5 mSv, and decreased with increased BMI (R2 = 0.45, p < 0.001). The mean ED to DAP ratio was 0.27 ± 0.04 mSv·Gy- 1·cm- 2 and tended to decrease with increased BMI (R2 = 0.55, p < 0.001). The mean ED to DAP ratios were 0.29 ± 0.02, 0.26 ± 0.02, and 0.23 ± 0.03 mSv·Gy- 1·cm- 2 for patients with BMI < 25 kg/m2, 25-30 kg/m2, and ≥30 kg/m2, respectively.
Conclusions:
Suitable conversion factors for C-arm CBCT facilitate the use of DAPs for estimating the ED. The patient dose can be varied by adjusting the CBCT rotation angle setting, and dose reduction strategies can be further manipulated.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Abdominal Aorta
The celiac trunk, a singular artery, divides into the left gastric artery, which...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...