Related Experiment Video
Updated: Feb 14, 2026

Analysis of HBV-Specific CD4 T-cell Responses and Identification of HLA-DR-Restricted CD4 T-Cell Epitopes Based on a Peptide Matrix
Published on: October 20, 2021
Task-shifting of CD4 T cell count monitoring by the touchscreen-based Muse™ Auto CD4/CD4% single-platform system for
André Kouabosso1, Christian Diamant Mossoro-Kpinde2, Ralph-Sydney Mboumba Bouassa3
1Centre National de Référence des Infection Sexuellement Transmissibles et de la Thérapie Antirétrovirale, Bangui, Central African Republic.
Background:
The accuracy of CD4 T cell monitoring by the recently developed flow cytometry-based CD4 T cell counting Muse™ Auto CD4/CD4% Assay analyzer (EMD Millipore Corporation, Merck Life Sciences, KGaA, Darmstadt, Germany) was evaluated in trained lay providers against laboratory technicians.
Methods:
After 2 days of training on the Muse™ Auto CD4/CD4% analyzer, EDTA-blood samples from 6 HIV-positive and 4 HIV-negative individuals were used for CD4 T cell counting in triplicate in parallel by 12 trained lay providers as compared to 10 lab technicians.
Results:
Mean number of CD4 T cells in absolute number was 829 ± 380 cells/μl by lay providers and 794 ± 409 cells/μl by technicians (P > 0.05); and in percentage 36.2 ± 14.8%CD4 by lay providers and 36.1 ± 15.0%CD4 by laboratory technician (P > 0.05). The unweighted linear regression and Passing-Bablok regression analyses on CD4 T cell results expressed in absolute count revealed moderate correlation between CD4 T cell counts obtained by lay providers and lab technicians. The mean absolute bias measured by Bland-Altman analysis between CD4 T cell/μl obtained by lay providers and lab technicians was -3.41 cells/μl. Intra-assay coefficient of variance (CV) of Muse™ Auto CD4/CD4% in absolute number was 10.1% by lay providers and 8.5% by lab technicians (P > 0.05), and in percentage 5.5% by lay providers and 4.4% by lab technicians (P > 0.05). The inter-assay CV of Muse™ Auto CD4/CD4% in absolute number was 13.4% by lay providers and 10.3% by lab technicians (P > 0.05), and in percentage 7.8% by lay providers and 6.9% by lab technicians (P > 0.05).
Conclusions:
The study demonstrates the feasibility of CD4 T cell counting using the alternative flow cytometer Muse™ Auto CD4/CD4% analyzer by trained lay providers and therefore the practical possibility of decentralization CD4 T cell counting to health community centers.
More Related Videos
Related Concept Videos
Design Example: Resistive Touchscreen
When a user touches the screen, the two layers make contact at a specific point known as the touchpoint. This contact reduces the resistance between...
Short-distance Transport of Resources
Numerical Calculations
The solution to a problem is obtained using different methods. While manually solving algebraic symbols is one of the most common methods, the graphical method is often preferred. Computers...
Pilot and Numeric Relaying
How Data are Classified: Numerical Data
Quantitative data may be either discrete or continuous. All quantitative data that take on only specific numerical...
Cancers Originate from Somatic Mutations in a Single Cell