Combination of the best pacing configuration and atrioventricular and interventricular delays optimization in cardiac
Pierre Socie1, Fabien Squara2, Marc Semichon1
1Department of Electrophysiology and Pacing, Centre Médico-Chirurgical Ambroise Paré, Neuilly-sur-Seine, France.
Background:
Cardiac resynchronization therapy optimization can be pursued by left ventricular pacing vector selection and atrioventricular (AV) and interventricular (VV) delays optimization. The combination of these methods and its comparison with multipoint pacing (MPP) is scarcely studied.
Methods:
Using noninvasive cardiac output (CO) measurement, the best of five left ventricular pacing vectors was determined, then AV and VV delays optimization was applied on top of the best vector. Response to the optimization protocol was defined as a >5% CO increase compared to the standard biventricular configuration.
Results:
Twenty-two patients (18 men, age 71 ± 9 years) were included. Standard biventricular configuration increased CO compared to baseline (4.65 ± 1.55 L/min vs 4.27 ± 1.53 L/min, respectively, P = 0.02). The best quadripolar configuration increased CO to 4.85 ± 1.67 L/min (P = 0.03 compared to the standard biventricular configuration). AV then VV delay optimization both provided additional benefit (final CO 5.56 ± 2.03 L/min, P = 0.001 compared to the best quadripolar configuration). Fifteen (68%) patients responded to the optimization protocol. Anatomical MPP (based on maximal anatomical separation between electrodes) and electrical MPP (based on maximal electrical activation difference between electrodes) were evaluated in 16 patients and yielded a CO similar to that of the optimization procedure.
Conclusions:
The combination of choosing the best quadripolar pacing configuration and optimizing atrioventricular and interventricular delays resulted in an improvement of cardiac output compared to standard biventricular stimulation in 68% of patients. The final cardiac output was comparable to multipoint pacing.
Related Concept Videos
Combination Therapies and Personalized Medicine
The combination of the drug acetazolamide and sulforaphane is a good example of combination therapy to treat cancer. The cells in the interior of a large tumor often die due to the hypoxic and...
Controller Configurations
Control-system compensation involves various configurations, most commonly series or cascade compensation, in which the controller...
Electron Configurations
The relative energies of the subshells determine the order in which atomic orbitals are filled (1s, 2s, 2p, 3s, 3p,...
Electron Configuration of Multielectron Atoms
Configurations of BJT
The common base configuration is noted for its high voltage gain, positioning it as an ideal choice for single-stage amplifier circuits, such as microphone pre-amplifiers. A notable...
Stability of Equilibrium Configuration
A stable equilibrium occurs when a system tends to return to its original position when given a small displacement, and the potential energy is at its minimum. An example of a stable equilibrium is when a cantilever beam is fixed at one end and a weight is attached to the other end. If the weight...