Related Experiment Video
Updated: Feb 14, 2026

The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
Left atrial appendage patency and device-related thrombus after percutaneous left atrial appendage occlusion: a
Hubert Cochet1,2, Xavier Iriart3, Soumaya Sridi1
1Department of Cardiovascular Imaging, Hôpital Cardiologique du Haut-Lévêque, CHU de Bordeaux, Avenue de Magellan, Bordeaux-Pessac, France.
Aims:
Transoesophageal echocardiography studies have reported frequent peri-device leaks and device-related thrombi (DRT) after percutaneous left atrial appendage (LAA) occlusion. We assessed the prevalence, characteristics and correlates of leaks and DRT on cardiac computed tomography (CT) after LAA occlusion.
Methods And Results:
Consecutive patients underwent cardiac CT before LAA occlusion to assess left atrial (LA) volume, LAA shape, and landing zone diameter. Follow-up CT was performed after >3 months to assess device implantation criteria, device leaks and DRT. CT findings were related to patient and device characteristics, as well as to outcome during follow-up. One-hundred and seventeen patients (age 74 ± 9, 37% women, CHA2DS2VASc 4.4 ± 1.3, and HASBLED 3.5 ± 1.0) were implanted with Amplatzer cardiac plug (ACP)/Amulet (71%) or Watchman (29%). LAA patency was detected in 44% on arterial phase CT images and 69% on venous phase images. The most common leak location was postero-inferior. LAA patency related to LA dilatation, left ventricular ejection fraction impairment, non-chicken wing LAA shape, large landing zone diameter, incomplete device lobe thrombosis, and disc/lobe misalignment in patients with ACP/Amulet. DRT were detected in 19 (16%), most being laminated and of antero-superior location. DRT did not relate to clinical or imaging characteristics nor to implantation criteria, but to total thrombosis of device lobe. Over a mean 13 months follow-up, stroke/transient ischaemic attack occurred in eight patients, unrelated to DRT or LAA patency.
Conclusion:
LAA patency on CT is common after LAA occlusion, particularly on venous phase images. Leaks relate to LA/LAA size at baseline, and device malposition and incomplete thrombosis at follow-up. DRT is also quite common but poorly predicted by patient and device-related factors.
Related Concept Videos
Surface Appendages of Archaea
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Cardiac Catheterization III: Left Heart Catheterization
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET