Related Experiment Video
Updated: Feb 14, 2026

Bioelectric Analyses of an Osseointegrated Intelligent Implant Design System for Amputees
Published on: July 15, 2009
In-vivo study of osseointegration in Prestige LP cervical disc prosthesis
Jigang Lou1, Beiyu Wang1, Tingkui Wu1
1Department of Orthopedics, West China Hospital, Sichuan University, 37 Guoxue Road, Chengdu, Sichuan, 610041, China.
Background:
A study was designed to quantify the extent of porous osseointegration at the prosthesis-bone interface in the Prestige LP prosthesis containing a plasma-sprayed titanium coating.
Methods:
Using an anterior surgical approach, cervical disc arthroplasty was performed in 8 mature male goats at the C3-C4 segment, followed by implantation of the Prestige LP prosthesis. The vertebral specimens were examined using microcomputed tomograph for histomorphometric quantification, and proceeded by routine paraffin processing for histological observation. Hence, the porous osseointegration at the prosthesis-bone interface was evaluated based on histologic and histomorphometric analyses.
Results:
At 6 months after surgery, there was no evidence of prosthesis migration, loosening, subsidence, or neurologic or vascular complications. Based on gross histologic analysis, there was excellent porous ingrowth at the prosthesis-bone interface, without significant histopathologic changes. Histomorphometric analysis at the prosthesis-bone interface indicated the mean porous ingrowth of 48.5% ± 10.4% and the total ingrowth range of 36.6 to 59.8%.
Conclusions:
As the first comprehensive in vivo investigation into the Prestige LP prosthesis, this project established a successful animal model in the evaluation of cervical disc arthroplasty. Moreover, histomorphometric analysis of porous ingrowth at the prosthesis-bone interface was more favorable for cervical disc arthroplasty with the Prestige LP prosthesis compared to historical reports of appendicular total joint arthroplasty.
Related Concept Videos
Design Example: Calculating Safe Diameter for Wind-Exposed Disc
Case Studies
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Equivalence: In Vitro and In Vivo Bioequivalence
Longitudinal Studies
Observational Studies
There are three types of observational studies – Prospective, retrospective, and cross-sectional.
Prospective Study
Prospective studies, also known as longitudinal or cohort studies, are carried out by collecting future data from groups sharing similar characteristics. One...