Comparison of four techniques for spine stereotactic body radiotherapy: Dosimetric and efficiency analysis
Saif Aljabab1, Balamurugan Vellayappan2, Eric Vandervoort1
1Radiation Medicine Program, The Ottawa Hospital Cancer Centre, Ottawa, ON, Canada.
Purpose:
The aim of this study is to compare the dosimetric differences between four techniques for spine stereotactic body radiotherapy (SBRT): CyberKnife (CK), volumetric modulated arc therapy (VMAT), and helical tomotherapy (HT) with dynamic jaws (HT-D) and fixed jaws (HT-F).
Materials/Methods:
Data from 10 patients were utilized. All patients were planned for 24 Gy in two fractions, with the primary objectives being: (a) restricting the maximum dose to the cord to ≤ 17 Gy and/or cauda equina to ≤ 20 Gy, and (b) to maximize the clinical target volume (CTV) to receive the prescribed dose. Treatment plans were generated by separate dosimetrists and then compared using velocity AI. Parameters of comparison include target volume coverage, conformity index (CI), gradient index (GI), homogeneity index (HI), treatment time (TT) per fraction, and monitor units (MU) per fraction.
Results:
PTV D2 and D5 were significantly higher for CK compared to VMAT, HT-F, and HT-D (P < 0.001). The average volume of CTV receiving the prescription dose (CTV D95) was significantly less for VMAT compared to CK, HT-F and HT-D (P = 0.036). CI improved for CK (0.69), HT-F (0.66), and HT-D (0.67) compared to VMAT (0.52) (P = 0.013). CK (41.86) had the largest HI compared to VMAT (26.99), HT-F (20.69), and HT-D (21.17) (P < 0.001). GI was significantly less for CK (3.96) compared to VMAT (6.76) (P = 0.001). Likewise, CK (62.4 min, 14059 MU) had the longest treatment time and MU per fraction compared to VMAT (8.5 min, 9764 MU), HT-F (13 min, 10822 MU), and HT-D (13.5 min, 11418 MU) (P < 0.001).
Conclusion:
Both CK and HT plans achieved conformal target coverage while respecting cord tolerance. Dose heterogeneity was significantly larger in CK. VMAT required the least treatment time and MU output, but had the least steep GI, CI, and target coverage.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits
Body Temperature