Effect of manual eyelid manipulation on intraocular pressure measurement by rebound tonometry
Sung Uk Baek1,2, Ahnul Ha1,2, Young Kook Kim1,2
1Department of Ophthalmology, Seoul National University College of Medicine, Seoul, Republic of Korea.
Background/Aims:
To investigate the effect of eyelid manipulation on the measurement of intraocular pressure (IOP) using two different tonometries (rebound tonometry (RT) vs Goldmann applanation tonometry (GAT)).
Methods:
103 patients with primary open-angle glaucoma were prospectively enrolled. For all of the patients, IOP measurements were performed in three different ways: (1) RT with lid manipulation (LM), (2) RT without LM and (3) GAT. The order of the three measurements was randomly selected. Additionally, the palpebral fissure height (PFH; elliptical space between upper and lower eyelids) was measured.
Results:
The mean value of IOP measured by GAT was 13.97±2.80 mm Hg, which was not significantly different from that by RT without LM (13.75±2.44 mm Hg; P=0.096), but which was significantly lower than that by RT with LM (15.21±2.91 mm Hg; P<0.001). On a Bland-Altman plot, RT with LM was overestimated relative to GAT (mean: -1.5) and RT without LM (mean: -1.2). Among the high IOPs (>20 mm Hg), interestingly, those measured by RT without LM were significantly lower than those measured by GAT (P<0.001). In the subgroup analysis of PFH, the smaller the PFH, the more exaggerated the IOP difference between GAT (P=0.014) and RT with LM (P<0.001).
Conclusion:
RT-measured IOP was significantly exaggerated when manipulation was applied to the eyelid. This overall trend was more pronounced when PFH was small. GAT-measured IOP, meanwhile, showed a good correlation with IOP measured using RT without LM.
Related Concept Videos
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Vapor Pressure