Related Experiment Video
Updated: Feb 14, 2026

Handheld Metal Detector Screening for Metallic Foreign Body Ingestion in Children
Published on: September 11, 2018
Improved Detection of Foreign Bodies on Radiographs Using X-ray Dark-Field and Phase-Contrast Imaging
Purpose:
The aim of this study was to investigate whether the detection of foreign bodies can be improved using dark-field and phase-contrast radiography compared with conventional (transmission) radiographs.
Materials And Methods:
Experiments were performed using ex vivo pig paws, which were prepared with differently sized foreign bodies of metal, wood, and glass (n = 10 each). Paws without foreign bodies served as controls (n = 30). All images were acquired using an experimental grating-based large object radiography system. Five blinded readers (second- to fourth-year radiology residents) were asked to assess the presence or absence of any foreign body. Sensitivity and specificity for the detection of metal, wood, glass, and any foreign body were calculated and compared using McNemar test and generalized linear mixed models.
Results:
Sensitivity for the detection of metal foreign bodies was 100% for all readers and image combinations. The sensitivity for the detection of wooden foreign bodies increased from 2% for transmission images to 78% when dark-field images were added (P < 0.0001). For glass foreign bodies, sensitivity increased from 84% for transmission images to 96% when adding phase-contrast images (P = 0.041). Sensitivity for the detection of any foreign body was 91% when transmission, dark-field, and phase-contrast images were viewed simultaneously, compared with 62% for transmission images alone (P < 0.0001). Specificity was 99% to 100% across all readers and radiography modalities.
Conclusions:
Adding dark-field images substantially improves the detection of wooden foreign bodies compared with the analysis of conventional (transmission) radiographs alone. Detection of glass foreign bodies was moderately improved when adding phase-contrast images.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
X-ray Imaging
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Dark Triad and Person Perception
Phase Diagrams