A simple distillation method to extract bromine from natural water and salt samples for isotope analysis by
1Équipe de Géochimie des Isotopes Stables, Sorbonne Paris Cité, Institut de Physique du Globe de Paris, UMR 7154, CNRS, 1 rue Jussieu, 75238, Paris Cedex 05, France.
Rationale:
In natural samples bromine is present in trace amounts, and measurement of stable Br isotopes necessitates its separation from the matrix. Most methods described previously need large samples or samples with high Br/Cl ratios. The use of metals as reagents, proposed in previous Br distillation methods, must be avoided for multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) analyses, because of risk of cross-contamination, since the instrument is also used to measure stable isotopes of metals.
Methods:
Dedicated to water and evaporite samples with low Br/Cl ratios, the proposed method is a simple distillation that separates bromide from chloride for isotopic analyses by MC-ICP-MS. It is based on the difference in oxidation potential between chloride and bromide in the presence of nitric acid. The sample is mixed with dilute (1:5) nitric acid in a distillation flask and heated over a candle flame for 10 min. The distillate (bromine) is trapped in an ammonia solution and reduced to bromide. Chloride is only distilled to a very small extent. The obtained solution can be measured directly by MC-ICP-MS for stable Br isotopes.
Results:
The method was tested for a variety of volumes, ammonia concentrations, pH values and distillation times and compared with the classic ion-exchange chromatography method. The method more efficiently separates Br from Cl, so that samples with lower Br/Cl ratios can be analysed, with Br isotope data in agreement with those obtained by previous methods.
Conclusions:
Unlike other Br extraction methods based on oxidation, the distillation method presented here does not use any metallic ion for redox reactions that could contaminate the mass spectrometer. It is efficient in separating Br from samples with low Br/Cl ratios. The method ensures reproducible recovery yields and a long-term reproducibility of ±0.11‰ (1 standard deviation). The distillation method was successfully applied to samples with low Br/Cl ratios and low Br amounts (down to 20 μg).
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometry: Isotope Effect
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Isotopes
An element's atomic mass, or weight,...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....