Quantitative urine test strip reading for leukocyte esterase and hemoglobin peroxidase
Matthijs N Oyaert1, Jonas Himpe1, Marijn M Speeckaert2
1Department of Laboratory Medicine, University Hospital Ghent, Ghent, Belgium.
Background:
Recently, urine test strip readers have become available for automated test strip analysis. We explored the possibilities of the Sysmex UC-3500 automated urine chemistry analyzer based on complementary metal oxide semiconductor (CMOS) sensor technology with regard to accuracy of leukocyte esterase and hemoglobin peroxidase results. We studied the influence of possible confounders on these measurements.
Methods:
Reflectance data of leukocyte esterase and hemoglobin peroxidase were measured using CMOS technology on the Sysmex UC-3500 automated urine chemistry analyzer. Analytical performance (imprecision, LOQ) as well as the correlation with white blood cell (WBC) and red blood cell (RBC) counts (Sysmex UF-5000) were studied. Furthermore, the influence of urinary dilution, haptoglobin, pH and ascorbic acid as confounders was determined.
Results:
Within- and between-run imprecision (reflectance signal) ranged from 1.1% to 3.6% and 0.9% to 4.2% for peroxidase and 0.4% to 2.5% and 0.4% to 3.3% for leukocyte esterase. Good agreement was obtained between the UF-5000 for RBCs and peroxidase reflectance (r=0.843) and for WBCs and leukocyte esterase (r=0.821). Specific esterase activity decreased for WBC counts exceeding 100 cells/μL. Haptoglobin influenced the peroxidase activity, whereas leukocyte esterase and peroxidase activities showed a pH optimum between 5.0 and 6.5. A sigmoidal correlation was observed between urinary osmolality and peroxidase activity.
Conclusions:
CMOS technology allows to obtain high quality test strip results for assessing WBC and RBC in urine. Quantitative peroxidase and leukocyte esterase are complementary with flow cytometry and have an added value in urinalysis, which may form a basis for expert system development.
Related Concept Videos
Hemoglobin
When all four heme groups are bound to oxygen, the resulting molecule is called oxyhemoglobin. As a result, arterial blood...
Filtration and Urine Formation
Uncertainty in Measurement: Reading Instruments
Urine Studies II: Urine Culture and Sensitivity Test
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Disorders of Leukocytes
Leukopenia may result from bone marrow disorders, autoimmune diseases, and infectious diseases. For example, conditions such as multiple myeloma and aplastic anemia can impair the bone marrow's ability to produce adequate leukocytes. Similarly, autoimmune diseases like lupus and viral infections such as HIV can prompt the immune...