Determination of Cardiac Output in Pulmonary Hypertension Using Impedance Cardiography
Marion Dupuis1, Elise Noel-Savina2, Gregoire Prévot2
1Department of Pulmonology, University Hospital of Toulouse, Toulouse, Francedupuis.m@chu-toulouse.fr.
Impedance cardiography (IPc) reliably measures cardiac output (CO) in pulmonary hypertension (PH) patients compared to reference methods. Accuracy may decrease for extreme CO values.
Area of Science:
- Cardiology
- Medical Devices
- Pulmonary Hypertension Research
Background:
- Cardiac output (CO) is a critical prognostic indicator in pulmonary hypertension (PH).
- Right heart catheterization (RHC) via direct Fick or thermodilution (TD) are gold standards for CO measurement.
- Impedance cardiography (IPc) offers a non-invasive alternative for CO assessment.
Purpose of the Study:
- To evaluate the accuracy of the PHYSIOFLOW® system for CO measurement in PH patients.
- To compare IPc-derived CO with reference methods (Fick and TD) in specific PH groups (1 and 4).
Main Methods:
- Prospective study involving 75 PH patients at a reference center.
- Simultaneous CO measurements using Fick, TD, and IPc during RHC.
- Bland-Altman analysis to assess agreement between CO measurement techniques.
Main Results:
- Mean CO values were similar across Fick (5.7±1.9 L/min), TD (5.4±1.5 L/min), and IPc (5.5±1.7 L/min).
- Low bias observed between IPc and Fick (0.149 L/min) and IPc and TD (-0.153 L/min).
- Correlation diminished at extreme CO levels (<3 L/min or >7 L/min); BMI and PH group influenced agreement.
Conclusions:
- IPc measurement of CO in PH patients demonstrates reliability comparable to direct Fick and TD methods.
- The accuracy of IPc for CO assessment may be reduced in cases of extreme cardiac output values.
More Related Videos
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Pulmonary Hypertension: Classification and Pathogenesis
There are various classifications for PH, each relating to different underlying causes and also...