Comparison of two intraoral scanners based on three-dimensional surface analysis
Related Experiment Video
Updated: Feb 14, 2026

Semi-Automated Planimetric Quantification of Dental Plaque Using an Intraoral Fluorescence Camera
Published on: January 27, 2023
1Department of Orthodontics, School of Dentistry, Chonnam National University, 33 Yongbong-ro, Buk-gu, Gwangju, 61186, South Korea. ortholkm@jnu.ac.kr.
Background:
This in vivo study evaluated the difference of two well-known intraoral scanners used in dentistry, namely iTero (Align Technology) and TRIOS (3Shape).
Methods:
Thirty-two participants underwent intraoral scans with TRIOS and iTero scanners, as well as conventional alginate impressions. The scans obtained with the two intraoral scanners were compared with each other and were also compared with the corresponding model scans by means of three-dimensional surface analysis. The average differences between the two intraoral scans on the surfaces were evaluated by color-mapping. The average differences in the three-dimensional direction between each intraoral scans and its corresponding model scan were calculated at all points on the surfaces.
Results:
The average differences between the two intraoral scanners were 0.057 mm at the maxilla and 0.069 mm at the mandible. Color histograms showed that local deviations between the two scanners occurred in the posterior area. As for difference in the three-dimensional direction, there was no statistically significant difference between two scanners.
Conclusions:
Although there were some deviations in visible inspection, there was no statistical significance between the two intraoral scanners.
Related Concept Videos
Dimensional Analysis
Conversion Factors and Dimensional Analysis
The unit...
Dimensional Analysis
In fluid mechanics, dimensional...
Dimensional Analysis
Dimensional analysis allows us to analyze and compare physical quantities on a...
Dimensional Analysis
Problem Solving: Dimensional Analysis
Three-Dimensional Analysis of Strain