Noise Exposure From High-Flow Nasal Cannula Oxygen Therapy: A Bench Study on Noise Reduction
Takamitsu Kubo1, Hiroaki Nakajima2, Ryo Shimoda2
1Medical Equipment Center, Shizuoka Cancer Center Hospital, Shizuoka Prefecture, Japan. t.kubo@scchr.jp.
Background:
High-flow nasal cannula (HFNC) oxygen therapy produces noise at a level such that patients often complain. However, the noise level has not been measured digitally.
Methods:
We evaluated 3 types of HFNCs without filters and 2 types with filters attached for noise reduction. Optiflow (with and without a filter), MaxVenturi (with and without a filter) and AIRVO2 (without a filter only) were positioned at the center of a hospital room. We measured the noise levels at the distance of 1 m from the equipment at various total flows (30, 40, 50, 60 L/min) and FIO (0.40, 0.60, and 0.90).
Results:
Noise levels were increased with the AIRVO2 and MaxVenturi when total flow and FIO were increased. Noise levels decreased with the MaxVenturi when a filter was used. The noise level did not change with the Optiflow when total flow and FIO were increased. The noise level decreased in the groups with AIRVO2 and Optiflow compared with MaxVenturi without a filter.
Conclusions:
The findings in this study show that the noise level of HFNC/Venturi could be reduced by attaching an intake filter. However, the noise level of HFNC/blender and HFNC/turbine decreased in comparison with HFNC/Venturi without an intake filter.
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Oxidation-Reduction Reactions
Nose and Nasal Cavity
Gene Therapy