Related Experiment Video
Updated: Feb 14, 2026

Animal Models of Depression - Chronic Despair Model CDM
Published on: September 23, 2021
Evaluation of the Novel Centrifugal Pump, CAPIOX SL, in Chronic Large Animal Experiments
Kei Iizuka1, Nobumasa Katagiri1, Yoshiaki Takewa1
1Department of Artificial Organs, National Cerebral and Cardiovascular Center Research Institute, Osaka, Japan.
Abstract:
In the development of a new device for extracorporeal circulation, long-term durability and biocompatibility are required. The CAPIOX SL Pump (SL pump, Terumo Corporation, Tokyo, Japan), which is a centrifugal pump using a two-pivot bearing, was developed with the hope of suppressing pump thrombus formation around the bearings. This study aimed to evaluate the in vivo performance of the SL pump in the condition assumed severe clinical situation for long-term extracorporeal membrane oxygenation (ECMO) support. Extracorporeal circulation using the SL pump was installed in three goats, with drainage from the inferior vena cava and infusion into the right jugular artery. The animals were maintained with target pump flow of 2.0-3.0 L/min for 3 or 7 days. Anticoagulation was performed by continuous infusion of heparin with a target activated coagulation time (ACT) of 200 ± 50 s. Blood tests were performed regularly. After 3 or 7 days, autopsies were performed on all animals. The pumps were disassembled and observed for thrombus formation. The results were compared with those of our previous study of the current model of the centrifugal pump (SP pump). All animals were successfully managed within target pump flows and ACT values during the scheduled period, with no adverse events. No thrombus formation was found around the bearing of the SL pump. The blood tests showed normal major organ functions, and platelet consumption and hemolysis were significantly lower in this study compared to the previous study of the SP pump. The CAPIOX SL Pump showed excellent durability and biocompatibility in a large animal experiment.
More Related Videos
Related Concept Videos
Centrifugation
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...
ATP Driven Pumps II: P-type Pumps
A typical P-type pump has three cytosolic domains: nucleotide-binding (N), phosphorylation (P), and activator (A) domains. These domains are connected to the membrane-spanning helices by short amino acid segments. ATP hydrolysis and covalent phosphoenzyme intermediate formation are crucial parts of the catalytic cycle. At the highly...
Centrifugal Force
Pumped Concrete
For direct-acting pumps, the concrete enters the pump via the inlet valve under the action of gravity and suction created by the movement of the piston. This concrete is then forced into the pipeline and out through the outlet valve by the forward movement...
Refrigerators and Heat Pumps
A household refrigerator removes heat from...