Effects of Pulsatile Blood Flow on Oxygenator Performance
Lotte Schraven1, Andreas Kaesler1, Christian Flege2
1Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, RWTH Aachen University, Aachen, Germany.
Abstract:
Extracorporeal membrane oxygenation (ECMO) is mainly used for the therapy of acute respiratory distress syndrome and chronic obstructive lung disease. In the last years, the development of these systems underwent huge steps in optimization, but there are still problems with thrombus formation, clogging, and thus insufficient gas exchange. One idea of ECMO optimization is a pulsatile blood flow through the oxygenator, but this is still a controversy discussion. Analyzing available publications, it was not possible to identify a general statement about the effect of pulsatile blood flow on the gas exchange performance. The variety of parameters and circuit components have such a high influence on the outcome that a direct comparison of the studies is difficult. For this reason, we performed a structured study to evaluate the effects of pulsatile blood flow on the gas exchange performance of oxygenator. In in vitro tests according to DIN EN ISO 7199, we tested a small oxygenator (0.25 m2 exchange surface, polymethylpentene fibers, 33 mL priming volume) with constant and pulsatile blood flow in comparison. Therefore, we varied the mean blood flow from 250 to 1200 mL/min, the amplitude of 0, 20, and 50%, and the frequency of 30, 60, and 90 bpm. The results demonstrate that the gas transfer for pulsatile and constant blood flow was similar (oxygen: 36-64 mLO2 /LBlood ; carbon dioxide 35-80 mLCO2 /LBlood ) for the same mean blood flow ranges. Over all, the results and analyses showed a statistically nonsignificant difference between pulsatile and nonpulsatile flow. Consequently, we conclude that the implementation of pulsatile blood flow has only a small to no effect on the gas exchange performance in an oxygenator. As the results were obtained using an oxygenator with a coiled fiber bundle, the test must be verified for a stacked fiber oxygenator.
More Related Videos
05:58Frequent Tail-tip Blood Sampling in Mice for the Assessment of Pulsatile Luteinizing Hormone Secretion
Published on: July 4, 2018
10:40Visualization and Analysis of Blood Flow and Oxygen Consumption in Hepatic Microcirculation: Application to an Acute Hepatitis Model
Published on: August 4, 2012
Related Concept Videos
Oxygen Transport in the Blood
Blood Flow
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
Applications of Integration to Find Blood Flow
Physiology of the Genitourinary System I: Renal Blood Flow and Glomerular Filtration
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...