Safety and Feasibility of a Novel Active Fixation Temporary Pacing Lead
Mark Webster, Sanjeevan Pasupati, Nigel Lever
1Green Lane Cardiovascular Service, Auckland City Hospital, Private Bag 92024, Victoria St. West, Auckland 1142, New Zealand. MWebster@adhb.govt.nz.
Objective:
This first-in-human study evaluated the safety and technical feasibility of the Tempo temporary cardiac pacing lead (BioTrace Medical), which includes a novel fixation mechanism and soft tip.
Background:
Complications of temporary pacing leads include dislodgment, arrhythmias, and ventricular perforation. Temporary pacing applications have increased with transcatheter aortic valve replacement (TAVR) growth, for rapid pacing during balloon valvuloplasty (BAV) and valve deployment, and for periprocedural bradyarrhythmia support.
Methods:
Eligible patients required temporary pacing for TAVR, BAV, or electrophysiology (EP) procedures. Transthoracic echocardiograms were obtained at baseline and 24 hours after lead removal. Safety was defined as freedom from pericardial effusion requiring intervention or evidence of tamponade. Technical feasibility involved successful intracardiac delivery and pace capture. Additional evaluations included pacing threshold (PCT), rapid pacing, dislodgment, or sustained ventricular arrhythmias. Follow-up was to 30 days.
Results:
Twenty-five patients (60% female; mean age, 64 ± 19 years) underwent 13 TAVRs (7 Sapien 3 valves [Edwards Lifesciences], 4 CoreValves [Medtronic], and 2 Lotus valves [Boston Scientific]), 11 EP procedures, and 1 BAV at two New Zealand centers from January 2016 to June 2016. Safety was met in all patients, with no device-related adverse events. Technical feasibility was achieved in 23 cases (92%); 2 patients had unsuitable anatomy. No patient had lead dislodgment or sustained ventricular arrhythmias, and the final procedural PCT was 0.7 ± 0.5 mA. Rapid pacing was successful in all cases. Five patients had successful postprocedural use up to 5 days.
Conclusion:
This first-in-human study demonstrates the safety and technical feasibility of the Tempo lead, providing stable periprocedural temporary pacing support.
Related Concept Videos
Survey Safety
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...