An ex vivo comparison of vascular access devices used in extracorporeal photopheresis
Related Experiment Video
Updated: Feb 14, 2026

Author Spotlight: Hickman Catheter Use for Long-Term Vascular Access in a Preclinical Swine Model
Published on: March 31, 2023
Jeffrey Szymanski1, Priyank Shah1, Marian Dynis1
1Department of Pathology and Immunology, Washington University in St. Louis, St. Louis, Missouri.
Background:
Central venous access devices are commonly used in extracorporeal photopheresis, but their performance has not been systematically evaluated. The primary objective of this study was to compare pressures at various flow rates for central venous access devices in an ex vivo simulation of photopheresis.
Study Design And Methods:
Diluted, heparinized red blood cells were circulated through central access devices in series with a photopheresis system, and pressures at several flow rates were recorded. The devices tested were the Trifusion catheter (Hickman), the Vortex single-lumen and dual-lumen ports (Angiodynamics), and the TidalPort device (Norfolk). Flow rates were also compared for silicone and polyurethane catheters and for different catheter internal diameters.
Results:
The Vortex dual-lumen port generated pressure alarms above flow rates of 60 mL/minute. Throughout flow rates from 5 to 100 mL/minute, the Trifusion catheter and the TidalPort device operated at lower pressures than the Vortex ports. Within typical clinical flow rates, neither catheter material nor internal diameter substantially affected pressure.
Conclusion:
Central venous access devices show large differences in pressure within flow rates used routinely in clinical settings. These differences cannot be fully attributed to catheter material composition or catheter internal diameter.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Seedless Vascular Plants
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis
Evolutionary Relationships through Genome Comparisons