Trefoil balloon for aortic valvuloplasty.

B Meier, B Friedli, I Oberhänsli

    British Heart Journal
    |September 1, 1986
    PubMed
    Summary

    A novel Trefoil balloon effectively treated congenital aortic stenosis in a pediatric patient. This innovative valvuloplasty device maintained blood flow during dilation, significantly reducing the aortic transvalvar gradient.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Coarctation of the aorta: pre and postoperative evaluation with MRI and MR angiography; correlation with echocardiography and surgery.

    The international journal of cardiovascular imaging·2005
    Same author

    Balloon aortic valvoplasty in paediatric patients: progressive aortic regurgitation is common.

    Heart (British Cardiac Society)·2003
    Same author

    [Clinical course following partial cavo-pulmonary bypass: the Geneva experience from 1985 to 2000].

    Archives des maladies du coeur et des vaisseaux·2003
    Same author

    [False aneurysm complicating ligation of patent ductus arteriosus operated on successfully].

    Annales de cardiologie et d'angeiologie·2003
    Same author

    Effects of cisapride on ventricular depolarization-repolarization and arrhythmia markers in infants.

    Biology of the neonate·2001
    Same author

    Transcatheter occlusion of the patent ductus arteriosus with a single device technique: comparison between the Cook detachable coil and the Rashkind umbrella device.

    International journal of cardiology·2001

    Area of Science:

    • Cardiology
    • Pediatric Cardiology
    • Interventional Cardiology

    Background:

    • Congenital aortic stenosis poses a significant risk in pediatric patients.
    • Traditional balloon valvuloplasty can lead to complications and suboptimal outcomes.
    • The development of innovative devices is crucial for improving treatment efficacy.

    Observation:

    • A 12-year-old boy with congenital aortic stenosis was treated using a new Trefoil balloon.
    • The Trefoil balloon features three angioplasty balloons on a single shaft, forming a rosette upon inflation.
    • This design allows for continuous blood flow through the aortic valve during the procedure.

    Findings:

    • Simultaneous inflation of the Trefoil balloon maintained blood flow, with aortic pressure dropping only to 75 mm Hg.
    • The balloon dilatation lasted 15 seconds at 4 bar (400 kPa).
    • The procedure successfully reduced the transvalvar gradient from 45 mm Hg to 15 mm Hg.

    Implications:

    • The Trefoil balloon represents a promising advancement in pediatric valvuloplasty techniques.
    • This device may offer improved safety and efficacy for treating congenital aortic stenosis.
    • Further research and clinical application of the Trefoil balloon could enhance outcomes for young patients with this condition.