Effect of gantry rotation speed and scan mode on peristalsis motion artifact frequency and severity at abdominal CT
Rutwik Shah1,2, Rhanna Khoram3, Jack W Lambert3
1Department of Radiology, University of California San Francisco, 505 Parnassus Avenue, San Francisco, CA, 94143-0628, USA. rutwik.shah@ucsf.edu.
Purpose:
The purpose of the study was to understand the effect of CT gantry speed and axial vs. helical scan mode on the frequency and severity of bowel peristalsis artifacts.
Method:
We retrospectively identified 150 oncologic abdominopelvic CT scans obtained on a 256 slice CT scanner: 50 scans obtained with Axial mode and 0.5-s gantry rotation time (Slow-Axial); 50 with Axial mode and 0.28-s gantry rotation time (Fast-Axial); and 50 scans with Helical mode and 0.28-s gantry rotation time (Fast-Helical). The patients included 74 women and 76 men with a mean age of 61 years (range 22-85 years). Two readers viewed all CT scans to record the presence and severity of bowel peristalsis artifact, location of artifact (stomach, duodenum/jejunum, ileum, and colon) and artifact location relative to bowel interface (gas-bowel, fluid-bowel, and gas-fluid). The severity of artifacts was recorded subjectively on a 3-point scale, and objectively based on maximum length of the artifact.
Results:
Peristalsis artifact was more commonly seen with Slow-Axial scan acquisition (37 of 50 patient scans, or 74%) than Fast-Axial (15 in 50 patient scans, or 30%, p < 0.001) and Fast-Helical (22 of 50 patient scans, or 44%, p < 0.005). The bowel segment distribution and severity of peristalsis artifacts were not significantly different between scan techniques.
Conclusion:
Peristalsis artifacts are common at abdominopelvic CT scans. Fast gantry rotation speed significantly reduces the frequency of bowel peristalsis artifacts and should be a consideration when imaging of bowel and structures near bowel is critical.
More Related Videos
07:20A Novel Approach to Overcome Movement Artifact When Using a Laser Speckle Contrast Imaging System for Alternating Speeds of Blood Microcirculation
Published on: August 30, 2017
07:51Measurements of Soil Carbon by Neutron-Gamma Analysis in Static and Scanning Modes
Published on: August 24, 2017
Related Concept Videos
Work and Power for Rotational Motion
Similarly, the power delivered to a system that is rotating about a fixed axis...
Rotational Motion about a Fixed Axis
Work-Energy Theorem for Rotational Motion
Relative Motion Analysis using Rotating Axes - Acceleration
Time differentiation is...
Relative Motion Analysis using Rotating Axes
However, to express the relative position of point B relative to point A, an additional frame of reference, denoted as x'y', is necessary. This additional frame not only translates but also rotates relative to the fixed frame, making it...
Equation of Motion: Rotation About a Fixed Axis
The tangential component is dependent on the direction of the angular acceleration of the flywheel. The tangential component of the acceleration propels the flywheel along its path. On the other hand,...