Magic Angle in Cardiac CT: Eliminating Clinically Relevant Metal Artifacts in Pacemaker Leads with a Lead-Tip/Gantry
Sebastian D Reinartz1, Christiane K Kuhl1, Kerstin Fehrenbacher1
1Department of Diagnostic and Interventional Radiology, University Hospital, RWTH Aachen University, Pauwels Strasse 30, 52074 Aachen, Germany.
Rationale And Objective:
To identify the influence of various parameters for reducing artifacts in computed tomography (CT) of commonly used pacemakers or implantable cardioverter-defibrillator (ICD) lead tips.
Materials And Methods:
This ex vivo phantom study compared two CT techniques (Dual-Energy CT [DECT] vs. Dual-Source CT [DSCT]), as well as the influence of incremental alterations of current-time product and pacemaker lead-tip angle with respect to the gantry plane. Four pacemaker leads and one ICD lead were evaluated. The images were assessed visually on a five-point Likert scale (1 = artifact free to 5 = massive artifacts). Likert values 1-3 represent clinically relevant, diagnostic image quality.
Results:
344 of 400 total images were rated with diagnostic image quality. The DECT and dual-source DSCT technique each scored 86% diagnostic image quality. Statistically, DECT images showed significantly improved image quality (P < .05). Concerning the current-time product, no statistically significant change was found. Regarding lead-tip positioning, an angle of ≤70° yielded 100% diagnostic image quality. Pacemaker and ICD leads were assessed to have statistically significant differences.
Conclusions:
Surprisingly, the lead-tip angle of 70° has been established as the key angle under which diagnostic image quality is always ensured, regardless of the imaging technique. Thus, we call 70° the "Magic angle" in CT pacemaker imaging.
Related Concept Videos
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle of Twist: Problem Solving
Phase-lead and Phase-lag Controllers
Design Example: Traverse Angle Computations
Angle Closure Glaucoma: Treatment