Measurement uncertainty of platelet concentration using the Sysmex XN automated hematology analyzer
Yong Kwan Lim1, Oh Joo Kweon1, Jee-Hye Choi1
1a Department of Laboratory Medicine , Chung-Ang University College of Medicine , Seoul , South Korea.
Abstract:
We estimated the measurement uncertainty (MU) of platelet concentration measured using the Sysmex XN system with two reference platelet counting methods described by DIN 58932-5 (PTB method) and the International Council for Standardization in Haematology (ICSH method). Ten blood samples were used to estimate and compare the MU of the XN system, and 30 samples were used to compare the methods. The standard uncertainty of the reference method was significantly higher for the ICSH method; the PTB method showed higher platelet concentrations than the ICSH method. When applying different methods with the XN system, optic counting showed higher MU compared to the other methods. There was good correlation among the two reference methods and three automated platelet-counting methods. We evaluated the MU in platelet concentrations measured using an automated hematology analyzer. Our results suggest that using the PTB method for calculating MU of the automated hematology analyzer is superior to the ICSH method because of its lower standard uncertainty.
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Uncertainty in Measurement: Significant Figures
Uncertainty in Measurement: Accuracy and Precision
The Uncertainty Principle
Drug Concentrations: Measurements
Plasma...
Uncertainty: Overview