Related Experiment Video
Updated: Feb 14, 2026

Fabrication of Flexible Image Sensor Based on Lateral NIPIN Phototransistors
Published on: June 23, 2018
Investigation of Sensitivities and Drift Effects of the Arrayed Flexible Chloride Sensor Based on RuO₂/GO at
Shi-Chang Tseng1, Tong-Yu Wu2, Jung-Chuan Chou3,4
1Graduate School of Mechanical Engineering, National Yunlin University of Science and Technology, Douliu 64002, Taiwan. tsengsc@yuntech.edu.tw.
Abstract:
We investigate the temperature effect on sensing characteristics and drift effect of an arrayed flexible ruthenium dioxide (RuO₂)/graphene oxide (GO) chloride sensor at different solution temperatures between 10 °C and 50 °C. The average sensor sensitivities according to our experimental results were 28.2 ± 1.4 mV/pCl (10 °C), 42.5 ± 2.0 mV/pCl (20 °C), 47.1 ± 1.8 mV/pCl (30 °C), 54.1 ± 2.01 mV/pCl (40 °C) and 46.6 ± 2.1 mV/pCl (50 °C). We found the drift effects of an arrayed flexible RuO₂/GO chloride sensor in a 1 M NaCl solution to be between 8.2 mV/h and 2.5 mV/h with solution temperatures from 10 °C to 50 °C.