Objective:
A new endovascular tool, the Liungman Guidewire Fixator, has been developed to simplify endovascular treatment in complex aortic aneurysms. The device has been extensively tested in bench models and animal trials. To verify the safety and functionality demonstrated in the porcine model, the device was tested in ten patients undergoing endovascular aortic repair (EVAR) or fenestrated endovascular aortic repair (f-EVAR) treatment for abdominal aortic aneurysm.
Methods:
The Liungman Guidewire Fixator consists of a braided stent-like, cylindrical structure with conical ends and a central channel for a 0.035″ guidewire. When in use, it is slid along the guidewire and positioned in the target artery, where the Liungman Guidewire Fixator interacts with the arterial wall by anchoring the guidewire to the wall through a radial force. The Liungman Guidewire Fixator allows for uninterrupted blood flow passed the point of fixation. In this study, the Liungman Guidewire Fixator was tested in ten patients undergoing EVAR or f-EVAR treatment for abdominal aortic aneurysm. The device was deployed and retrieved crossover into the hypogastric artery, and the occurrence of thrombotic occlusion, arterial dissection, and vascular rupture or trauma was studied using angiography, as well as device ability to withstand guidewire tension.
Results:
There were no instances of occlusion, dissection, or vascular trauma detected using angiography. In all cases, deployment and retrieval were successful, and the devices could withstand an applied tension of 3 N. In one instance, retrieval was challenging because of significant tortuosity, which was resolved by a coaxial catheterization.
Conclusions:
Deployment was uneventful in all ten patients. Retrieval according to the intended instruction for use was performed in nine of the patients. In one patient, a coaxial catheterization was necessary. All devices withstood a retention force of 3 N.
Related Concept Videos
Survey Safety
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Household Wiring And Electrical Safety
Carbon-dioxide Fixation
Nursing Clinical Information System
A Nursing Clinical Information System (NCIS) is a specialized type of healthcare information system tailored to meet the unique needs of nursing practice. It incorporates the principles of nursing informatics to streamline information management and improve the quality of care delivery.
Critical attributes of NCIS include:
Clinical Trials
There are four phases in a clinical trial. A phase one...