A Biomechanical Comparison of Fifth Metatarsal Jones Fracture Fixation Methods
Neil L Duplantier1, Ronald J Mitchell1, Steve Zambrano2
1Houston Methodist Hospital, Orthopedics & Sports Medicine, Houston, Texas, USA.
Background:
Fifth metatarsal base fractures of the metaphyseal-diaphyseal watershed junction (Jones fracture) are commonly treated with surgical fixation in athletes. Intramedullary screw fixation remains the most utilized construct, although plantar-lateral plating is an alternative. Purpose/Hypothesis: The purpose was to compare the mechanical strength of fracture fixation between an intramedullary screw and plantar-lateral plating. The hypothesis was that plantar-lateral plate fixation would allow for more cycles and higher peak loads before failure, as well as less fracture gapping, than would an intramedullary screw in cadaveric foot specimens with simulated Jones fractures exposed to cantilever bending.
Study Design:
Controlled laboratory study.
Methods:
Twelve pairs of male cadaver feet were separated into 2 groups (plate or screw) to conduct contralateral comparative testing of 2 devices with equally numbered right and left feet. For each fifth metatarsal, an osteotomy with a microsagittal saw was created to simulate a Jones fracture. The plate group underwent fixation with a 3.0-mm 4-hole low-profile titanium plate placed plantar-laterally with 3 locking screws and 1 nonlocking screw. The screw group underwent fixation with a 40- or 45-mm × 5.5-mm partially threaded solid titanium intramedullary screw. After fixation, the metatarsals were excised for biomechanical testing. Cyclic cantilever failure testing was conducted with a gradient-cycle method. Sinusoidal loading forces were applied, increasing by 5.0-pound-force increments per 10 cycles, until each specimen experienced mechanical failure of implant or bone. Failure mode, number of cycles to failure, peak failure load, gap width at the last mutual prefailure loading, and video data were recorded. Paired 2-tailed t test (α = 0.05) was used to compare groups ( P < .05 set for significance).
Results:
Failure mode in both groups occurred predominantly at the bone-implant interface. Plate fixation resulted in significantly higher mean ± SD values for cycles to failure (63.9 ± 27.0 vs 37.3 ± 36.9, P = .01) and peak failure load (159.2 ± 60.5 N vs 96.5 ± 45.8 N, P = .01), with a significantly lower mean gap width (0.0 ± 0.0 mm vs 3.2 ± 2.4 mm, P < .01).
Conclusion:
As compared with intramedullary screw fixation, plantar-lateral plating allowed for greater cycles to failure and peak load before failure, as well as less gap width, when applied to cadaver foot specimens with simulated Jones fractures exposed to cantilever bending in a load frame.
Clinical Relevance:
Early return to play among athletes before Jones fracture union is associated with increased risk of failure. This study introduces a plantar-lateral plating construct that performed more favorably than intramedullary screw fixation when applied to simulated Jones fractures in cadaveric foot specimens. Further clinical comparative studies are needed to assess the clinical use of this construct.
More Related Videos
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
Related Concept Videos
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Carbon-dioxide Fixation
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...