Improving Quality of Dynamic Airway Computed Tomography Using an Expiratory Airflow Indicator Device
Lewis D Hahn1, Arthur W Sung2, Majid Shafiq3
1Stanford University Department of Radiology, Division of Thoracic Imaging.
Purpose:
Dynamic computed tomography (CT) of the airways is increasingly used to evaluate patients with suspected expiratory central airway collapse, but current protocols are susceptible to inadequate exhalation caused by variable patient compliance with breathing instructions during the expiratory phase. We developed and tested a low-cost single-use expiratory airflow indicator device that was designed to improve study quality by providing a visual indicator to both patient and operator when adequate expiratory flow was attained.
Materials And Methods:
A total of 56 patients undergoing dynamic airway CT were evaluated, 35 of whom were scanned before introduction of the indicator device (control group), with the rest comprising the intervention group. Lung volumes and tracheal cross-sectional areas on inspiratory/expiratory phases were computed using automated lung segmentation and quantitative software analysis. Inadequate exhalation was defined as absolute volume change of <500 mL during the expiratory phase.
Results:
Fewer patients in the intervention group demonstrated inadequate exhalation. The average change in volume was higher in the intervention group (P=0.004), whereas the average minimum tracheal cross-sectional area was lower (P=0.01).
Conclusions:
The described expiratory airflow indicator device can be used to ensure adequate exhalation during the expiratory phase of dynamic airway CT. A higher frequency of adequate exhalation may improve reliability and sensitivity of dynamic airway CT for diagnosis of expiratory central airway collapse.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Indicators
Imaging Studies III: Computed Tomography
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...