Updated: Feb 14, 2026

Measurement of Tumor T2* Relaxation Times after Iron Oxide Nanoparticle Administration
Published on: May 19, 2023
Measuring liver T2* and cardiac T2* in a single acquisition
Suraj D Serai1,2, Andrew T Trout3, Robert J Fleck3
1Department of Radiology, MLC 5031, Cincinnati Children's Hospital Medical Center, 3333 Burnet Avenue, Cincinnati, OH, 45229, USA. serais@email.chop.edu.
Purpose:
The purpose of this study is determine if both liver T2* and cardiac T2* can be measured on a single breath-hold acquisition.
Materials And Methods:
For this IRB-approved retrospective study, 137 patients with dedicated Cardiac MRI and Liver MRI examinations obtained sequentially on 1.5T scanners and on the same day were included for analysis. Both the cardiac and liver MRI examinations utilized GRE sequences for quantification of tissue iron. Specifically, T2* was measured using an 8-echo, multi-echo gradient echo single breath-hold sequence. Liver T2* was measured in a blinded manner on images from each of the cardiac and dedicated liver MRI examinations and were correlated. Bland-Altman difference plot was used to assess mean bias.
Results:
137 examinations from 93 subjects met inclusion criteria. 10 examination pairs were excluded because the first echo time (TE) on the cardiac MRI was insufficiently short for the very high liver iron content. After exclusion, 127 studies from 89 subjects (67.4% males) were included in the final analysis. The mean subject age (± standard deviation) was 11.5 ± 7.5 years (range 0-29.3 years; median 10.5 years). Mean liver T2* measured on cardiac MRI was 8.3 ± 7.7 ms and mean liver T2* measured on dedicated liver MRI was 7.8 ± 7.4 ms (p < 0.001). There was strong positive correlation between the two liver T2* measurements (r = 0.989, p < 0.0001; 95% CI 0.985-0.992). With the exception of borderline outliers, all values fell within two standard deviations on the Bland-Altman difference plots, with a mean bias of 0.5 ms (range - 1.8 to + 2.7 ms).
Conclusion:
In most patients with suspected or known iron overload, a single breath-hold GRE sequence may be sufficient to evaluate the iron concentration (T2*) of both the myocardium and the liver.
More Related Videos
Related Concept Videos
Light Acquisition
Water and Mineral Acquisition
Liver Histology
Hepatocytes perform a variety of essential functions. They secrete...
Liver Physiology
Metabolic Regulation:
The liver is the central organ involved in regulating blood composition. It stabilizes blood glucose levels, maintaining them within the range of 70–110 mg/dL. When these levels drop, the liver breaks down glycogen reserves and releases glucose into the bloodstream. It can...
Liver Regeneration
Cells of Liver
The liver comprises four major types of cells— hepatocytes, stellate, Kupffer, and sinusoidal endothelial cells. The hepatocytes are...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...