Related Experiment Video
Updated: Feb 14, 2026

An In Vitro Bladder Model of Catheter-Associated Urinary Tract Infection
Published on: June 24, 2025
Cluster-Randomized Trial of Devices to Prevent Catheter-Related Bloodstream Infection
Steven M Brunelli1, David B Van Wyck2, Levi Njord2
1DaVita Clinical Research, Minneapolis, Minnesota.
Insights
ClearGuard HD antimicrobial barrier caps significantly reduced bloodstream infections (BSIs) in dialysis patients using central venous catheters (CVCs). This advancement offers improved safety and care for hemodialysis patients at risk of infection.
Area of Science:
- Nephrology
- Infectious Disease
- Medical Devices
Background:
- Central venous catheters (CVCs) are a major source of bloodstream infections (BSIs) in dialysis patients, leading to increased hospitalizations, mortality, and costs.
- Existing interventions may reduce BSIs, but comparative-effectiveness data is limited, hindering evidence-based clinical decisions for dialysis access.
Purpose of the Study:
- To compare the effectiveness of ClearGuard HD antimicrobial barrier caps versus Tego hemodialysis connectors with Curos disinfecting caps in reducing BSI rates.
- To provide robust data for informed patient care decisions regarding CVC infection prevention in hemodialysis.
Main Methods:
- A 13-month, prospective, cluster-randomized, open-label trial involving 40 US DaVita dialysis facilities.
- Facilities were pair-matched and randomized 1:1 to either ClearGuard HD caps or Tego+Curos caps.
- 1671 adult dialysis patients with CVCs were enrolled; the primary outcome was the positive blood culture (PBC) rate indicating BSI.
Main Results:
- Baseline BSI rates were similar between groups before intervention (P=0.8).
- During the intervention period, the ClearGuard HD group showed a significantly lower BSI rate (0.28 PBCs/1000 CVC-days) compared to the Tego+Curos group (0.75 PBCs/1000 CVC-days; P=0.001).
- No device-related adverse events were reported during the study.
Conclusions:
- ClearGuard HD antimicrobial barrier caps significantly reduced catheter-related BSIs in hemodialysis patients compared to Tego connectors with Curos caps.
- This finding represents a significant advancement in preventing infections associated with CVCs in hemodialysis care.
Abstract:
Central venous catheters (CVCs) contribute disproportionately to bloodstream infection (BSI) and, by extension, to infection-related hospitalization, mortality, and health care costs in patients undergoing dialysis. Recent product advancements may reduce BSIs, but a sufficiently powered comparative-effectiveness study is needed to facilitate evidence-based patient care decisions. In a 13-month, prospective, cluster-randomized, open-label trial, we compared BSI rates in facilities using ClearGuard HD antimicrobial barrier caps (ClearGuard group) with those in facilities using Tego hemodialysis connectors plus Curos disinfecting caps (Tego+Curos group). Forty DaVita dialysis facilities in the United States were pair-matched by BSI rate, number of patients using CVCs, and geographic location, and then cluster randomized 1:1. We enrolled all adult patients undergoing dialysis with CVCs at these facilities, except those allergic to heparin or chlorhexidine. Overall, 1671 patients participated in the study, accruing >183,000 CVC-days. The study outcome was positive blood culture (PBC) rate as an indicator of BSI rate. We calculated results at the cluster level and adjusted for the facility cluster effect. During a 3-month run-in period immediately before study interventions, the groups had similar BSI rates (P=0.8). During the 13-month intervention period that immediately followed, the ClearGuard group had a BSI rate significantly lower than that of the Tego+Curos group (0.28 versus 0.75 PBCs per 1000 CVC-days, respectively; P=0.001). No device-related adverse events were reported. In conclusion, compared with Tego connectors plus Curos caps, ClearGuard HD antimicrobial barrier caps significantly lowered the rate of catheter-related BSIs in patients undergoing hemodialysis using CVCs, representing an important advancement in hemodialysis patient care.
Related Concept Videos
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Trial and Error and Algorithm
Clinical Trials: Overview
Clinical Trials
There are four phases in a clinical trial. A phase one...
Random Error
Random Variables
Uppercase letters such as X or Y denote a random variable. Lowercase letters like x or y denote the value of a random variable. If X is a random variable, then X is written in words, and x is given as a number.
For example, let X = the...