Comparison of Corneal Tomography: Repeatability, Precision, Misalignment, Mean Elevation, and Mean Pachymetry
Related Experiment Video
Updated: Feb 14, 2026

Corneal Donor Tissue Preparation for Endothelial Keratoplasty
Published on: June 12, 2012
Simon Schröder1, Stephanie Mäurer1, Timo Eppig1
1a Institute of Experimental Ophthalmology , Saarland University , Homburg/Saar , Germany.
Purpose:
To compare corneal tomography and its statistical uncertainty for measurements obtained by three clinically used corneal tomographers: A Scheimpflug camera (Pentacam HR), a swept source optical coherence tomography system (CASIA SS-1000), and Placido ring imaging (TMS-5).
Material And Methods:
Repeated measurements with all three devices on 34 normal eyes were used to estimate the repeatability, precision, and mean values of corneal elevation and pachymetry within 8 mm diameter. The repeatability (standard deviation) was calculated for each data point of the corneal elevation data-maps of anterior and posterior cornea as well as for the pachymetry data-maps. Uncertainty on the position of the eye at each measurement might contribute to the differences between elevation data-maps. To take this into account, we defined the precision as the standard deviation for the elevation data-maps of anterior and posterior cornea after correction of misalignment-effects (rotation, translation). The mean elevation and pachymetry data-maps were fitted with Zernike polynomials for interdevice-comparison.
Results:
Pentacam HR offered the best repeatability and precision for the anterior corneal elevation (<3 and <1.6 μm, respectively). CASIA SS-1000 offered good repeatability and precision with high resolution for posterior corneal elevation, and the best repeatability for pachymetry (<3 μm). TMS-5 measured anterior elevation with similar repeatability to CASIA SS-1000 (<6 μm). The data-maps of the three tomographers could not be used interchangeably. The largest differences were observed for pachymetry and posterior corneal elevation data-maps.
Conclusions:
Misalignment limited the repeatability of TMS-5 and Pentacam HR, but had little influence on the repeatability of CASIA SS-1000.
Related Concept Videos
Freezing Point Depression and Boiling Point Elevation
The boiling point of a liquid is the temperature at which its vapor pressure is equal to ambient atmospheric pressure. Since the vapor pressure of a solution is lowered due to the presence of nonvolatile solutes, it stands to reason that the solution’s boiling point will subsequently be increased. Vapor pressure increases with temperature, and so a solution will require a higher temperature than will pure solvent to achieve any given vapor pressure, including one...
Uncertainty in Measurement: Accuracy and Precision
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Elevation of Intermediate Points on Vertical Curves
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Accuracy and Precision