Improved diagnostic accuracy of thallium-201 myocardial perfusion single-photon emission computed tomography with CT
Jei-Yie Huang1,2, Ruoh-Fang Yen1, Wen-Chung Lee2
1Department of Nuclear Medicine, National Taiwan University Hospital and National Taiwan University College of Medicine, No. 7, Chung Shan S. Rd, Taipei City, 10002, Taiwan.
Background:
The benefits of attenuation correction (AC) in technetium-99m myocardial perfusion imaging (MPI) have been well established. However, the value of thallium (Tl-201) AC and routine computed tomography AC (CTAC) were less well established. The aims of this study were to evaluate the diagnostic performance of thallium (Tl-201) MPI with additional CTAC and to determine which participants would benefit most.
Methods And Results:
A total of 108 consecutive patients who underwent Tl-201 MPI and received coronary angiography within 3 months were enrolled. Diagnostic performance was determined by sensitivity, specificity, and receiver operating characteristic curve analysis. Subgroup analyses were performed using gender and obesity. CTAC improved the area under the curve (0.84 vs. 0.77, P = 0.037 at patient level), primarily due to a significant improvement in specificity (0.78 vs. 0.57, P = 0.013) and no significant difference in sensitivity (0.79 vs. 0.82, P = 0.75). In subgroup analysis, CTAC was most helpful in obese subjects, men, and especially right coronary artery lesions.
Conclusions:
CTAC significantly improved diagnostic performance primarily by increasing the specificity, and the improvements were significantly greater in obese patients and male patients. These findings suggest that CTAC should be applied to Tl-201 MPI as routine clinical practice.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Improving Translational Accuracy
Improving Translational Accuracy
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies III: Computed Tomography
Emission Spectra