A Comparison of Surface Infrared with Rectal Thermometry in Dogs
T O Omóbòwálé1, B N Ogunro, E A Odigie
1Department of Veterinary Medicine, University of Ibadan, Ibadan. bukitayo_omobowale@yahoo.com.
Abstract:
Accurate determination of temperature is crucial in the diagnosis of febrile conditions. Although fewer techniques have proven as useful and reliable a predictor of core body temperature as the rectal thermometry, the process of obtaining the rectal temperature could be stressful in dogs. The infrared thermometry is a noncontact device used for measuring body temperature, with advantages which include speed, convenience, and reduced stress to the animals and reduced occupational risks to the animal handler. Therefore, there is the need to assess the consistency and agreement between non-contact infrared thermometry and traditional rectal thermometry in body temperature estimation. This study compared and assessed the sensitivity of non-contact infrared thermometer used on the forehead and nasal regions respectively with that of a rectal thermometer in dogs for body temperature estimation. One hundred and thirty (130) dogs presented for veterinary attention at the Veterinary Teaching Hospital (VTH), University of Ibadan, Nigeria were enrolled in this study during August to September 2014, irrespective of sex, age, breed or health status. Temperatures of dogs presented at the clinic were obtained using both multiple non-contact infrared thermometric measures obtained in the nasal and frontal head regions; and by rectal temperature. A multivariate cross-matrix analysis was used to assess the difference in measurements between the rectal thermometry and non-contact infrared thermometry. Descriptive statistics was used to compare variation and trend regularity of the nasal and fore-head infrared thermometry. A logistic regression of the difference in measurements was computed at 95% confidence interval and P<0.05. The mean difference revealed that the rectal temperature was 5.330C higher than the non-contact infrared forehead-based temperature and 7.570C higher than nasal-based temperature measurements respectively. The Bland-Altman (B-A) plot showed that the 95% limits of agreement between the frontal and nasal obtained infrared laser thermometry methods. Temperature measure obtained using non-contact infrared thermometry (forehead and nasal region of the head) was poor in consistency and agreement compared to rectal thermometry. Usefulness of non-contact forehead infrared thermometry in routine clinical practice as a close estimate of core body temperature depends on accurate calibration and therefore not recommended.
Related Concept Videos
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits